![natus Echo-Screen III Pro Instructions For Use Manual Download Page 33](http://html1.mh-extra.com/html/natus/echo-screen-iii-pro/echo-screen-iii-pro_instructions-for-use-manual_3504910033.webp)
© 2020 Natus Medical Incorporated
29
026057 RevE
Testing Equipment
Echo-Screen III Pro Instructions for Use
A green check mark indicates that the cable is functioning properly. A red “X” indicates that
the test failed.
6
If an Out of Bounds error occurred, make sure each clip is seated properly in the test fixture
and try again. If the test continues to fail, contact Natus Technical Service or your authorized
service representative.
Hardware
Test the hardware:
1
Disconnect all cables from the device.
2
On the Home screen, touch
Equipment Check
.
3
On the Equipment Check screen, touch
Systems Test
.
4
On the Systems Test screen, touch
Hardware Test
. The test begins immediately.
5
Check the test result and then touch
Done
.
A green check mark indicates that the test passed. A red “X” indicates that the test failed.
6
If the hardware test failed, contact Natus Technical Service or your authorized service
representative.