Preparing for Testing
Echo-Screen III Pro Instructions for Use
Create To-Do Lists for later screenings:
•
Add a patient to the To-Do List:
•
On the Home screen, touch
To-Do List
. Touch
Add
, enter the patient information, and
touch
Save
.
•
On the Home screen, touch
Patient List
. Select a patient already on the device, and
then touch
To-Do
.
•
Remove a patient from the To-Do list:
•
On the Home screen, touch
To-Do List
. Select a patient in the list, and then touch
Remove
.
Note:
The patient is removed from the To-Do List but remains in the Patient
List.
Preparing for Testing
IMPORTANT:
Do not operate the Echo-Screen III Pro device close to display monitors or other
electrical equipment. Before you begin, check the calibration date displayed on the Home
screen. You cannot perform a screening when the probe is out of calibration.
WARNING
Do not use the Echo-Screen III device to screen or connect to a patient while it is
docked in the docking station or plugged into external power.
Do not turn off medical equipment suspected of causing interference unless under
the direct supervision of a physician or appropriate medical practitioner.
If you are permitted to use the Echo-Screen III Pro integrated hooks on a bassinet,
hang the device at the foot of the bassinet, well away from the baby.
Never leave a patient unattended while the patient is connected to the device.
Do not suspend the Echo-Screen III device or the wireless printer over the infant’s
head.
WARNING
If using the wireless printer with the shoulder strap, remove or secure the printer
prior to preparing the patient.
CAUTION
Do not operate the Echo-Screen III Pro device in an environment where it continu-
ously displays interference. Such interference may slow down or stop the screening
and compromise the device’s performance or accuracy.
Do not operate the Echo-Screen III Pro device if cardiac monitors or incubator fans
(or other equipment required for monitoring a very sick infant) in the near vicinity
cause interference.