EN-10
EN-11
EN
EN
Measurement was cancelled prematurely. Repeat the meas
-
urement and wait until the measurement is compete before
removing the thermometer from the measurement point.
Calibration error. If this occurs repeatedly, please contact
Natural Cycles customer support.
The device is in technical test mode. Switch it off and on
again after a few seconds to return to normal mode.
Device repairs may only be performed by the manufacturer or an expressly
authorised body. Please contact:
We recommend a technical inspection of the device every 5 years. This in
-
spection can be carried out either by UEBE Medical GmbH or by authorised
maintenance services.
This product complies with Council Directive 93/42/EEC of 5 Sep
-
tember 2007 concerning medical devices and bears the mark CE
0123 (TÜV SÜD Product Service GmbH).
Degree of protection against electric shock TYPE BF
Consult the instructions for use
Manufacturer
Dust tight, protection against temporary submersion
NaturalCycles Nordic AB
St Eriksgatan 63b
112 34 Stockholm, Sweden
help.naturalcycles.com
www.naturalcycles.com
TECHNICAL DATA
CUSTOMER SERVICE
EXPLANATION OF SYMBOLS
Type:
Maximum thermometer, direct mode
Measuring points/body refer
-
ence points:
rectal, oral, vaginal
Temperature unit:
°C
Measuring range:
32.00°C to 43.99°C
Measurement accuracy:
± 0.05°C between 35.00°C and 38.00°C,
± 0.10°C in other temperature ranges
Memory:
30 measurement values with date and
time can be recalled in memory mode
Operating conditions:
Ambient temperature 10°C to 40°C,
maximum relative humidity 15% to
95%
Conditions for storage and
transport:
-25°C to +55°C, maximum relative
humidity 15% to 95%
Time to adjust from minimum/
maximum storage temperature
to operating temperature:
30 minutes
Battery:
1 x CR2032, 3.0 V, mercury-free
Battery life:
approx. 400 measurements
Data transfer:
Bluetooth 4.0 LE (low energy)
IP classification:
IP67: dust tight, protection against
temporary submersion
Expected service life:
5 years
Dimensions:
137 mm (L) x 32 mm (W) x 15 mm (H)
Weight:
27 g with battery
0123
IP 67
Sensor error. If this occurs repeatedly, please contact Natural
Cycles customer support.
Lot number/Batch number
Reference number = item number