background image

 

g.  Do not let the front lens element come into contact with a wet slide surface, as prolonged contact with any 

moisture could damage the lens.  If lens is exposed to moisture, promptly wipe with soft tissue to remove 
moisture. 

 
11.  Diopter adjustment for models 132 and 132-CLED dual head. 
 

These dual head models have a 30

°

 inclined eyepiece to provide comfortable viewing for the primary viewer.  They 

also have a vertical eyepiece so that a second viewer can observe the image simultaneously.   The primary viewer will 
have focused the microscope using the coarse and fine focus knobs.  To compensate for differences in vision 
between primary and second viewer, second viewer may have to adjust image in vertical eyepiece by turning the 
knurled diopter adjustment. 

 

MAINTENANCE 

 

WARNING:  For your own safety, turn switch off and remove plug from power source before maintaining your 
microscope.  If the power cord is worn, cut or damaged in any way, have it replaced immediately to avoid 
shock or fire hazard. 

 
1.  OPTICAL MAINTENANCE 
 

Do not attempt to disassemble any lens components.  Consult a microscope service technician when any repairs not 
covered by instructions are needed. 
 
Prior to cleaning any lens surface, brush dust or dirt off lens surfaces using a camel hair brush.  Or use air to blow 
dust and lint off surfaces.  Use of compressed air in a can,  available at any computer supply store, is a good source 
of clean air. 
 
Do not remove eyepieces or objective lenses to clean.  Clean only the outer lens surface.  Breath on lens to dampen 
surface, then wipe with lens paper or tissue or use a cotton swab moistened with distilled water.  Wipe lenses with a 
circular motion, applying as little pressure as possible.  Avoid wiping dry lens surface as lenses are scratched easily.  
If excessive dirt or grease gets on lens surfaces, a small amount of Windex or distilled water can be used on a cotton 
swab or lens tissue.  To clean objective lenses, do not remove objectives from microscope.  Clean front lens element 
only, following same procedure. 
 
NOTE:  Fingerprints or other matter on the front lens element of the objective lens is the single most common reason 
that you will have difficulty in focusing the microscope.  Before having costly servicing done, or before returning to 
National for “warranty repair”, make certain to examine the front lens element with a magnifying glass or eye loupe for 
the presence of such contaminants.  If a microscope is returned to National for warranty repair, and it is determined 
that such contaminants are the problem, this is not covered under warranty and National  will submit a cost estimate 
for cleaning. 

 
2.  MECHANICAL MAINTENANCE 
 

a.  The rack stop screw has been pre-adjusted at the factory and should not require re-adjustment.  However, if you 

do attempt re-adjustment, note the following procedure. 

 

Loosen round knurled locking nut by turning counter-clockwise.  Then loosen stop screw.  With fine focus 
adjustment at mid-range, focus on standard slide until sharp image is obtained.  Rotate rack stop screw in 
clockwise direction until tight.  Lock into position with the locking nut.  A needle nose pliers might be needed to 
loosen the round locking nut. 

 
b.  Coarse focus tension adjustment prevents the stage from drifting down from its own weight and causing the 

image to move out of focus. This has been adjusted at the factory, but over the course of time it may loosen and 
cause the stage of the microscope to slip downward on the focusing block. 

 

The tension adjustment collar is located between arm and coarse focus knob on left side of microscope.  With a 
small jewelers screwdriver (models 131/132-CLED use supplied 0.90 “L” hex wrench), loosen the set screw 
located in only one of the four holes on tension adjustment collar. Turn collar clockwise to tighten tension, 
counter-clockwise to loosen tension. Use of a wide rubberband will provide a better grip on the tension adjustment 
collar.  After adjusting, tighten the set screw to lock collar in place. 

 

Summary of Contents for 131

Page 1: ... These models share common features nomenclature and method of operation The only difference are Models 131 132 feature electric illumination from a tungsten bulb Models 131 CLED 132 CLED feature electric illumination from a LED bulb Models 132 132 CLED have both 30 inclined and vertical eyepiece tubes each supplied with WF10x eyepiece Copyright 1 2 01 National Optical Scientific Instrument Inc ...

Page 2: ...STAGE CLIPS Two locked on clips hold specimen slide in place on stage Note Your microscope is already drilled and tapped to accept an optional mechanical stage Mechanical stage replaces stage clips and permits precise mechanical manipulation of the specimen slide 6 STAGE Platform of the microscope where the specimen slide is placed 7 CONDENSER LENS An 0 65 N A condenser lens fixed in center of sta...

Page 3: ...therwise you will not be able to see anything when looking through the microscope b Standard lenses provided with your microscope are a widefield 10x eyepiece two eyepieces supplied with models 132 and 132 CLED 4x 10x and 40x objectives The 40x objective lens has a special spring retractable mechanism which retracts slightly if front of lens comes in contact with the specimen slide See chart below...

Page 4: ... swab or lens tissue To clean objective lenses do not remove objectives from microscope Clean front lens element only following same procedure NOTE Fingerprints or other matter on the front lens element of the objective lens is the single most common reason that you will have difficulty in focusing the microscope Before having costly servicing done or before returning to National for warranty repa...

Page 5: ...tly Replace with a 110v 20 watt lamp with double contact medium bayonet base This is a standard commonly used replacement bulb National 800 102 Carefully wipe lamp to insure it is clean and free of all fingerprints Replace field lens housing assembly b Changing LED bulb on Models 131 CLED 132 CLED To open the illuminators field lens housing use 0 9mm L hex key wrench supplied with your microscope ...

Page 6: ...n also result from dirt on inside of eyepiece It is recommended that you have service technician clean inside of lens OPTIONAL ACCESSORIES AND PARTS 610 045R WF10x eyepiece w reticle 10mm 100div 615 045 WF15x eyepiece w pointer 760 045 DIN 60xR objective lens N A 0 90 800 001 Replacement LED bulb 131 CLED 132 CLED only 800 102 Replacement bulb 115v 20 watt tungsten D C medium base 131 132 only 801...

Reviews: