2.  Insert microscope plug into matched voltage outlet.  On Models 131, 131-CLED & 132, flip on/off power switch located 

on base to “ON” position.  On Model 132-CLED, rotate on/off power switch located on base to “ON” position. 

 
3.  Rotate coarse focus knobs to move stage down (away) from objective lens as far as possible. 
 
4.  Place specimen slide, cover slip facing up, on stage with specimen centered over condenser lens in middle of stage. 
 
5.  Rotate disc diaphragm until largest aperture is positioned beneath condenser in center of stage. 
 
6.  Turn the objective turret until the 4x (smallest) objective lens “clicks” into position in the optical path.  Note that each 

time you change from one objective lens to another you should turn the turret until you hear the “click”, which 
indicates that the lens is properly indexed in the optical path. 

 
7.  While looking through the eyepiece, rotate coarse focusing knobs until specimen comes into focus. If image does not 

appear in field of view, move specimen slide slightly until image appears in field of view. 

 
8.  Adjust fine focus controls until specimen is in sharp focus. 
 
9.  Turn the disc diaphragm, observing that different apertures affect the sharpness of the image.  Turn diaphragm until 

sharpest possible image is obtained.  Note that when turning the disc diaphragm, you should hear a “click” as each 
aperture comes into proper position under the condenser lens.  If aperture is not properly positioned, you will observe 
shadows in the field of view when looking through the microscope. 

 
10.  Changing magnification 
 

a.  Magnification is changed by rotating objective turret until different objective lens is moved into optical path.  

Always turn turret until you hear the “click”, indicating that lens is properly indexed.  Otherwise, you will not be 
able to see anything when looking through the microscope. 

 
b.  Standard lenses provided with your microscope are a widefield 10x eyepiece (two eyepieces supplied with 

models 132 and 132-CLED). 4x. 10x, and 40x objectives.  The 40x objective lens has a special spring retractable 
mechanism which retracts slightly if front of lens comes in contact with the specimen slide. See chart below for 
specifications on objectives. 

 

Objective Specification Chart

 

 

Objective 

N.A. 

Color Code 

Ring 

Field of 

View 

Working 

Distance 

Magnification with 

WF10X eyepiece 

Din 4X 

0.10 

Red 

4.5mm 

18.5mm 

40X 

Din 10X 

0.25 

Yellow 

1.8mm 

6.4mm 

100X 

Din 40X retractable 

0.65 

Blue 

0.45mm 

0.5mm 

400X 

 

c.  Also note that each objective has a color ring, which permits you to instruct changes in magnification by referring 

to an easily observed color rather than to a number. 

 

d.  The microscope has been parfocalled at the factory, which allows easy change from one magnification to another, 

requiring little or no adjustment of the fine focus knobs. 

 
e.  As magnification is increased, the field of view (area of specimen seen through the microscope) will decrease.  

That is why it is easier to find the specific area of interest on the specimen by starting with the lowest 4x objective 
lens, before increasing magnification with the 10x or 40x objective lens. 

 
f. 

NOTE:  Care must be taken when rotating the 40x objective into place.  This lens has a spring retractable 
mechanism which retracts slightly into its housing if the front of the lens strikes the specimen slide.  With fine 
focus adjustment at mid-range, the rack stop has been adjusted at the factory to assure the 40x lens will clear the 
thickness of a normal specimen slide and cover slip.  However, if the rack stop has been improperly adjusted, or if 
using a thicker than normal slide or cover slip, moving the 40x lens too quickly or carelessly could cause damage 
to the front lens element or to the slide. 

 

Summary of Contents for 131

Page 1: ... These models share common features nomenclature and method of operation The only difference are Models 131 132 feature electric illumination from a tungsten bulb Models 131 CLED 132 CLED feature electric illumination from a LED bulb Models 132 132 CLED have both 30 inclined and vertical eyepiece tubes each supplied with WF10x eyepiece Copyright 1 2 01 National Optical Scientific Instrument Inc ...

Page 2: ...STAGE CLIPS Two locked on clips hold specimen slide in place on stage Note Your microscope is already drilled and tapped to accept an optional mechanical stage Mechanical stage replaces stage clips and permits precise mechanical manipulation of the specimen slide 6 STAGE Platform of the microscope where the specimen slide is placed 7 CONDENSER LENS An 0 65 N A condenser lens fixed in center of sta...

Page 3: ...therwise you will not be able to see anything when looking through the microscope b Standard lenses provided with your microscope are a widefield 10x eyepiece two eyepieces supplied with models 132 and 132 CLED 4x 10x and 40x objectives The 40x objective lens has a special spring retractable mechanism which retracts slightly if front of lens comes in contact with the specimen slide See chart below...

Page 4: ... swab or lens tissue To clean objective lenses do not remove objectives from microscope Clean front lens element only following same procedure NOTE Fingerprints or other matter on the front lens element of the objective lens is the single most common reason that you will have difficulty in focusing the microscope Before having costly servicing done or before returning to National for warranty repa...

Page 5: ...tly Replace with a 110v 20 watt lamp with double contact medium bayonet base This is a standard commonly used replacement bulb National 800 102 Carefully wipe lamp to insure it is clean and free of all fingerprints Replace field lens housing assembly b Changing LED bulb on Models 131 CLED 132 CLED To open the illuminators field lens housing use 0 9mm L hex key wrench supplied with your microscope ...

Page 6: ...n also result from dirt on inside of eyepiece It is recommended that you have service technician clean inside of lens OPTIONAL ACCESSORIES AND PARTS 610 045R WF10x eyepiece w reticle 10mm 100div 615 045 WF15x eyepiece w pointer 760 045 DIN 60xR objective lens N A 0 90 800 001 Replacement LED bulb 131 CLED 132 CLED only 800 102 Replacement bulb 115v 20 watt tungsten D C medium base 131 132 only 801...

Reviews: