NOTE:  It is recommended that you leave the tension as loose as possible for ease of focusing, yet not so loose 
that it permits the stage of microscope to drift downward from its own weight and cause the microscope to “drift” 
out of focus. 

 
c.  Metal parts:  Use a clean, damp cloth to remove dust or dirt from metal parts, followed by a dry cloth. 

 
 
3.  ELECTRICAL MAINTENANCE  
 

The extent of electrical maintenance, by other than a qualified technician, should be bulb replacement.  BE CERTAIN 
TO TURN SWITCHES OFF AND REMOVE PLUG FROM POWER SOURCE OUTLET BEFORE CHANGING 
BULBS. 

 

a.  Changing bulb on Models 131 & 132 

 

Make sure that illuminator field lens housing and lamp are cool before attempting to remove.  Remove illuminator 
field lens housing by rotating in a counter-clockwise direction.  This housing is secured tightly, to prevent easy 
removal by students, so a very firm grip and some strength will be required. 

 

Note that lamp does not screw out.  Remove lamp by depressing lamp slightly and rotate in a counter-clockwise 
direction until it pops up slightly.  Replace with a 110v 20 watt lamp with double contact, medium bayonet base.  
This is a standard, commonly used replacement bulb.  National #800-102.  Carefully wipe lamp to insure it is 
clean and free of all fingerprints.  Replace field lens housing assembly. 

 

b.  Changing LED bulb on Models 131-CLED & 132-CLED 

 

To open the illuminators field lens housing, use 0.9mm “L” hex key wrench supplied with your microscope.  
Loosen hex screws on lens housing.  Remove lens housing to expose LED bulb.  Remove bulb by grasping the 
plastic base of bulb and gently pulling straight up.  Insert new LED bulb, replace lens housing and tighten hex 
screws to secure lens housing in place. 
 

c.  Replacement of fuse on Models 131-CLED & 132-CLED 

 

The fuse is located at the rear right side of microscope base.  To remove fuse from holder, insert a 6mm 
screwdriver blade into slot located in rear of fuse holder cap.  Slightly depress and rotate screwdriver ¼ turn in 
direction of arrow, release pressure on screwdriver to release fuse.  Pull cap and fuse out of fuse holder.  Insert 
proper fuse into fuse cap.  Insert fuse cap into fuse holder.  Using screwdriver, rotate fuse cap assembly in 
opposite direction of arrow until slot engages, depress fuse cap and rotate ¼ turn to lock into fuse holder. 
 

 

 

Summary of Contents for 131

Page 1: ... These models share common features nomenclature and method of operation The only difference are Models 131 132 feature electric illumination from a tungsten bulb Models 131 CLED 132 CLED feature electric illumination from a LED bulb Models 132 132 CLED have both 30 inclined and vertical eyepiece tubes each supplied with WF10x eyepiece Copyright 1 2 01 National Optical Scientific Instrument Inc ...

Page 2: ...STAGE CLIPS Two locked on clips hold specimen slide in place on stage Note Your microscope is already drilled and tapped to accept an optional mechanical stage Mechanical stage replaces stage clips and permits precise mechanical manipulation of the specimen slide 6 STAGE Platform of the microscope where the specimen slide is placed 7 CONDENSER LENS An 0 65 N A condenser lens fixed in center of sta...

Page 3: ...therwise you will not be able to see anything when looking through the microscope b Standard lenses provided with your microscope are a widefield 10x eyepiece two eyepieces supplied with models 132 and 132 CLED 4x 10x and 40x objectives The 40x objective lens has a special spring retractable mechanism which retracts slightly if front of lens comes in contact with the specimen slide See chart below...

Page 4: ... swab or lens tissue To clean objective lenses do not remove objectives from microscope Clean front lens element only following same procedure NOTE Fingerprints or other matter on the front lens element of the objective lens is the single most common reason that you will have difficulty in focusing the microscope Before having costly servicing done or before returning to National for warranty repa...

Page 5: ...tly Replace with a 110v 20 watt lamp with double contact medium bayonet base This is a standard commonly used replacement bulb National 800 102 Carefully wipe lamp to insure it is clean and free of all fingerprints Replace field lens housing assembly b Changing LED bulb on Models 131 CLED 132 CLED To open the illuminators field lens housing use 0 9mm L hex key wrench supplied with your microscope ...

Page 6: ...n also result from dirt on inside of eyepiece It is recommended that you have service technician clean inside of lens OPTIONAL ACCESSORIES AND PARTS 610 045R WF10x eyepiece w reticle 10mm 100div 615 045 WF15x eyepiece w pointer 760 045 DIN 60xR objective lens N A 0 90 800 001 Replacement LED bulb 131 CLED 132 CLED only 800 102 Replacement bulb 115v 20 watt tungsten D C medium base 131 132 only 801...

Reviews: