14

 

 

 

5.2.2

 

CONNECTION TO THE ELECTRICITY MAINS 

The properties of the electricity mains (size, protection system, earth connection) must comply with the legislation 
applying to the electrical properties of the chiller in the country in which it is installed.  

The general regulations pertaining to installation are listed below. 

 

 

The electrical installation may only be carried out by qualified personnel! 

 

The mains connection should be established with a rubber-sheathed cable connected to a TN-C system with direct 

earthing.

  

 

230 V version (1 phase):  

Connection  with  a  single-phase,  three-wire  system  (1/N/PE  AC).  The  phase  conductor  and  the  neutral  conductor 

must be connected directly to the main switch of the chiller. 

 

The protective earth conductor must be connected to the earthing point of the chiller. 

 

400 V version (3 phase):   

Connection  with  a  three-phase,  four-wire  system  (3/PEN  AC).  The  three-phase  conductor  must  be  connected  di-

rectly to the main switch of the chiller. No neutral conductor is used. The PEN must be connected to the earthing 

point of the chiller.

 

If different mains voltages and frequencies are available, the country-specific features must be observed. 

 

 

Make sure that the mains voltage and the frequency match the data on the identification plate of the chiller and 
agree with those in the table in the supplement and that the permitted values are not exceeded. Check the mains 

for correct values and test the function of the earthing and protection systems. 

 

Install a general isolating switch for the electricity mains close to the chiller. The connecting circuit for the chiller 

must be equipped with a circuit breaker and residual current protective device.

 

  

You can find information on the electrical power input values for the chiller in the supplement. 

Before you connect the chiller to the mains, make sure that the general isolating switch disconnects the chiller from 
the electricity mains.  

Pass the wires in the flexible polymer cable gland in the control panel (see photo for position) to connect the power 

cable. Feed the connecting lead through the aperture in the rear plate (PCGE 31-PCGE 141) on the base of the 

chiller. Lay the cable on the upright up to the gland on the base of the switch cabinet (PCGE 200- PCGE 3300). In-

side the switch cabinet lay the cable in the installation duct up to the main switch, where it is connected directly. 

Check that the screws on the terminal clamps are firmly seated! 

 

   

    

 

 

 

 

 
 

 

 

 

 

 

 
 

 

 

 

 

 

 

 

Holes for feeding through the power cable 

Cable inlet 

Summary of Contents for PCGE Series

Page 1: ...00 OCGE 31 OCGE 240 PCGEZ 08 PCGEZ 480 English Operating and maintenance manual Deutsch Bedienungs und Wartungsanleitung National Lab GmbH Grambeker Weg 157 D 23879 Moelln Phone 49 0 45 42 84 91 10 Fa...

Page 2: ...RS 17 6 2 1 OIL REQUIREMENTS 17 6 2 2 CHECKLISTE FOR OCGE 31 TO OCGE 240 17 6 3 CHECK LIST FOR COMMISSIONING GLYCOL RECIRCULATING CHILLERS 18 6 3 1 REQUIREMENTS FOR THE WATER GLYCOL MIXTURE 18 6 3 2 C...

Page 3: ...y of cooling medium e g water water glycol moisture or oil constantly within certain temperature limits operating range with the aid of a technical cooling process The most commonly used cooling mediu...

Page 4: ...in this connection The side panels of the chiller must not be removed unless this is absolutely necessary Even if the side panels are only partially removed isolate the chiller from the electricity su...

Page 5: ...f fully hermetic design with integrated bimetal switch and oil sight glass Semi hermetic compressor Semi hermetic compressor with integrated bimetal switch and oil sight glass Reciprocating piston com...

Page 6: ...gs such as non return valves filter drier sight glass and service connections 3 2 2 OPTIONS REFRIGERATION CIRCUIT Condenser air filter aluminium reusable The condenser air filter protects the condense...

Page 7: ...return circuit systems with single and dual circuit The digital thermometer shows the inlet temperature of the cooling medium VNR Non return valve in the supply line These valves are used in open syst...

Page 8: ...ite failure report summarizes all fault messages from the compressor the pump s fan s of the high pressure and low pressure pressostats the flow monitor as well as the breakage of probes Voltage free...

Page 9: ...r by means of codes 3 4 2 OPTIONS ELECTRIC CIRCUIT Connecting cable with CEE plug A ready to use mains cable is supplied with 6m oilflex cable and CEE plug 4 pole 3P PEN or 5 pole 3P N PE Compressor w...

Page 10: ...GE 81 111 760 760 1330 1250 767 100 972 100 84 1445 1 PCGEZ 100 760 760 1330 1250 767 100 972 100 84 1445 1 PCGE 141 OCGE 141 PCGEZ 160 760 760 1380 1300 767 100 972 100 84 1510 1 MODELS PCGE 150 to P...

Page 11: ...r lifting it in accordance with the applicable safety standards If the chiller is not fitted with eyebolts the lifting gear must be appropriate for the weight of the chiller as indi cated on the ident...

Page 12: ...ting it against direct sunlight You should avoid connecting the axial ventilators to air ducts as they do not have sufficient compression Higher ambient temperatures than those given as a reference va...

Page 13: ...allow use of the consumer circuit if you have to remove the chiller from the circuit switch it off or use it in different ways at the same time A pipeline system with integrated filter near the chille...

Page 14: ...atures must be observed Make sure that the mains voltage and the frequency match the data on the identification plate of the chiller and agree with those in the table in the supplement and that the pe...

Page 15: ...plate on the motor 6 1 2 CHECKLIST FOR PCGE 11 TO PCGE 3300 Controller Type A see 6 4 PCGE 11 to PCGE 28 1 Remove the lid panel and the two side panels 2 Fill the collection tank Open the tank cap and...

Page 16: ...ches MTC and MTP are in position 0 8 Fill the tank to the maximum upper arrow level indicator is located on the bottom right of the frame See Section 6 1 1 Water requirements 9 Move the motor protecti...

Page 17: ...re required by the customer 26 When the chiller is within the operating conditions set by the manufacturer check whether the current uptake of the chiller agrees with the data indicated on the identif...

Page 18: ...dinary maintenance work each chiller start should be regarded as an initial start up The initial start up calls for a number additional checks to ensure safe and reliable performance of the system 6 3...

Page 19: ...cation plate and those contained in the table in the supplement Controller Type C see 6 4 PCGEZ 240 to PCGEZ 480 1 Remove the front plate and the two side panels 2 Open the tank cap and fill the tank...

Page 20: ...hes AUX Second relay active ON A temperature alarm has been triggered ON Type B PCGE 31 to 520 OCGE 31 to 240 PCGEZ 37 to 180 Symbols degrees Celsius degrees Fahrenheit bar PSI Compressor 1 Compressor...

Page 21: ...of the chiller listed in the supplement The operating thermostat controls both the temperature adaptation functions and the other safety parameters e g alarm management and the running time of the el...

Page 22: ...the parameter you require 6 Press the key to change its value 7 Change the value with the and keys 8 Press the key again to confirm the new value After a few seconds the next parameter will be display...

Page 23: ...ressed for 10 seconds 2 Move the main switch of the chiller to position 0 Via the main switch Move the main switch to position 0 The system then switches off immediately This procedure should only be...

Page 24: ...G FOR TRANSPORT PURPOSES All work on the chiller may only be conducted by specialist personnel Follow this procedure if you need to shut down the chiller Stop the chiller and permanently disconnect it...

Page 25: ...1 Check whether the current consumption of the compressor group is within the target values indicated in these instructions 2 Clean the fin pack of the condenser using a compressed air jet with a pre...

Page 26: ...constituents must be treated in line with the provisions of the valid legislation 9 MALFUNCTIONS TROUBLESHOOTING All service and maintenance operations may only be carried out by specialized personne...

Page 27: ...tion and in case change probe HA High liquid tem perature alarm Compressor switched off by internal protection Contact a technician 2 Pump is not working fan and com pressor are working Pump is blocke...

Page 28: ...low Fill the tank A09 Pump thermal switches alarm Thermal protections of a new unit are in OFF posi tion Switch MTP in ON M after reset MTP Pump is having problem Check pump and possible obstruc tion...

Page 29: ...probe is defec tive Probe can be defective or not well connected Check connection and in case change probe A 3 Unit is work ing A13 Maintenance circuit 1 only on request Service on unit is required du...

Page 30: ...Pump thermal switches alarm Thermal protections of a new unit are in OFF posi tion Switch MTP in ON M after reset MTP Pump is having problem Reset thermal switch and in case contact of a persistent p...

Page 31: ...tches alarm Thermal protections of a new unit are in OFF po sition Switch MTC in ON M after reset MTC Compressor is having problem Reset thermal switch and in case contact of a persistent problem con...

Page 32: ...ME VON OEL RUCKKUHLERN 47 6 3 1 ANFORDERUNGEN AN DAS OEL 47 6 3 2 CHECKLISTE OCGE 31 BIS OCGE 240 47 6 4 CHECKLISTE ZUR INBETRIEBNAHME VON GLYKOL RUCKKUHLERN 48 6 4 1 ANFORDERUNGEN AN DAS WASSER GLYKO...

Page 33: ...Menge Kal tetrager z B Wasser Wasser Glykol Gemisch oder Oel mit Hilfe eines technischen Kuhlverfahrens konstant inner halb bestimmter Temperaturgrenzen Betriebsbereich zu halten Der am haufigsten ver...

Page 34: ...gung des Herstellers durchgefuhrt werden wenn Komponenten des Ge rates unter Druck stehen Verschutten Sie niemals Kaltetrager da die Gefahr von gesundheitlichen Schaden besteht Vor einer eventuel len...

Page 35: ...G Das Kuhlgerat ist bei Lieferung anschlussfertig Es besteht aus einem Gehause einem vollstandig hermetisch abge dichteten Kaltemittelkreislauf dem Kaltetrager Kreislauf und der Elektrotechnik inklusi...

Page 36: ...ialventilator zum Anschluss an einen bauseitigen Luftkanal Einspritzverfahren a Kapillareinspritzung bis zu Kuhlleistungen von ca 2 5 kW Das Kaltemittel wird uber eine definierte Kapillare eingespritz...

Page 37: ...Der Schwimmerschalter befindet sich im Tank und uberwacht den Kaltetragerstand Beim Sinken des Wasserstandes und Erreichen des Mindestgrenzwertes werden das Kuhlsystem und die Pumpe n abgeschaltet Das...

Page 38: ...Einkreis System ohne Tank ausgelegt Die Pumpe fo rdert den Kaltetrager innerhalb des Kreislaufs durch den Verdampfer und uber die Verbraucher Der Kaltetrager wird im Verdampfer abgekuhlt Der Kreislau...

Page 39: ...ltet QE Schaltschrank Im Schaltschrank sind alle elektrische Komponenten und Sicherheitsvorrichtungen enthalten TLT Betriebsthermostat Die Steuerung und Regelung der Temperatur erfolgt durch den Betri...

Page 40: ...E 81 111 760 760 1330 1250 767 100 972 100 84 1445 1 PCGEZ 100 760 760 1330 1250 767 100 972 100 84 1445 1 PCGE 141 OCGE 141 PCGEZ 160 760 760 1380 1300 767 100 972 100 84 1510 1 MODELLE PCGE 150 bis...

Page 41: ...a den geltenden Sicherheits standards zum Heben des Gerats verwendet werden Wenn das Kuhlgerat nicht mit Hebeo sen ausgestattet ist muss das Hebemittel fur das auf dem Typenschild und im Anhang angege...

Page 42: ...trachtigen empfehlen wir es vordirekter Sonneneinstrahlung zu schutzen Ein Anschluss der Axial Ventilatoren an Luftkanale sollte vermieden werden da sie keine ausreichende Pressung aufweisen Ho here R...

Page 43: ...hern zu vermeiden Die empfohlene Maschenweite fur die Kuhlgerate liegt bei 25 mit einem maximalen Druckverlust am Filter von 1 bar Die Verwendung ei nes Filters mit einer Maschenweite von 60 eignet si...

Page 44: ...t Bevor Sie das Kuhlgerat mit dem Stromnetz verbinden stellen Sie sicher dass der allgemeine Trennschalter die Anschlussleitung des Kuhlgerates vom Stromnetz trennt Fuhren Sie die Drahte durch die fle...

Page 45: ...den vom Hersteller eingestellten Betriebsbedingungen befindet ob die elektrische Absorption des Kuhlgerates mit den auf dem Typenschild angegebenen und in der Tabelle im Beiblatt enthaltenen Daten ube...

Page 46: ...ebsbedingungen erreicht hat diese sind dann erreicht wenn das Display den ein gestellten Arbeitsbereich anzeigt 17 Uberprufen Sie ob sich der angezeigte Druck innerhalb der fur die Pumpe spezifizierte...

Page 47: ...nden OCGE Serie Tank in der Nahe des Kuhlgerates max 2m Zahnradpumpe Option mit weiter entferntem Tank 2m Oel Austrittstemperatur ISO32 min 20 C ISO46 min 25 C ISO68 min 30 C ISO10 ISO68 Allgemeine Be...

Page 48: ...n Hersteller fur diesen Zweck ausdrucklich zugelassen sind Der Glykolanteil muss mindestens 20 betragen und sollte einen Wert von 25 nicht uberschreiten Ka ltetra geraustrittstemperatur Ethylenglykol...

Page 49: ...die beiden Seitenbleche 2 O ffnen Sie die den Tankdeckel und befullen Sie den Tank ggf mit einem T Stuck unter Kontrolle der Fullstandsanzeige siehe dazu Abb 1 Punkt 5 2 1 3 O ffnen Sie die Beluftungs...

Page 50: ...EIN Ein Temperaturalarm ist aufgetreten EIN Typ B PCGE 31 bis 520 OCGE 31 bis 240 PCGEZ 37 bis 180 Symbole Grad Celsius Grad Fahrenheit Bar Psi Kompressor 1 Kompressor 2 Einheit im Standby Allgemeine...

Page 51: ...eiblatt aufgefuhrten Standardspezifikationen des Kuhlgerates in vor eingestelltem Zustand geliefert Der Betriebsthermostat steuert sowohl die Temperaturanpassungsfunktionen als auch andere Sicherheits...

Page 52: ...ppe von Parametern bedeutet 2 Benutzen Sie die und Pfeile um durch die Namen der zu andernden Parametergruppen zu scrollen 3 Nachdem Sie die richtige Parametergruppe gefunden haben drucken Sie die Tas...

Page 53: ...ie den Wert mit den und Tasten 6 Drucken Sie die Taste erneut um den neuen Wert zu bestatigen Nach einigen Sekunden wird der nachste Parameter erscheinen 7 Um das Programm zu verlassen Drucken Sie die...

Page 54: ...hrt werden Wenn Sie das Kuhlgerat abschalten mussen gehen Sie wie folgt vor Stoppen Sie das Kuhlgerat und trennen Sie es dauerhaft vom Stromnetz Erst nachdem Sie es ordnungsgema vom Netz getrennt habe...

Page 55: ...eckungen der Maschinenteile Wo chentlich 1 Prufen Sie von au en ob der Stand des Kaltetragers im Tank innerhalb der Grenzen liegt die auf der An zeige an der Seite des Kuhlgerates zu sehen ist 2 Prufe...

Page 56: ...halter die Stromzufuhr zum Kuhlgerat unterbricht und bringen Sie entsprechende Hinweisschilder an um eine erneute Verbindung der Maschine zum Stromnetz zu vermeiden Zur richtigen Handhabung oder Lager...

Page 57: ...erschmutzter Verflussiger Verflussiger wie in der Wartungs anleitung beschrieben reinigen Ventilator arbeitet nicht einwandfrei oder die Dreh richtung stimmt nicht Uberprufen Sie die Sicherung des Ven...

Page 58: ...en Sie ggf den Kundendienst Luftfilter verschmutzt nur bei Geraten die uber diese Option verfugen Demontieren Sie den Luftfilter vom Gerat und reinigen Sie ihn mit Druckluft oder Wasser LA Zu niedrige...

Page 59: ...ms erset zen Sie die Steuereinheit M 2 Ventilator und Kompressor funkti onieren nicht Pumpe funktioniert A01 Hochdruck Alarm beim Kaltemittel Zu hohe Umgebungs temperatur Umgebungstemperatur senken Be...

Page 60: ...Fuhler ist defekt Fuhler moglicherweise defekt oder nicht richtig angeschlossen Anschluss uberprufen und ggf Fuhler ersetzen A 3 Gerat funktioniert A13 Wartung von Kreislauf 1 nur bei Bedarf Wartung d...

Page 61: ...schal ter Pumpe Alarm der Pumpe Der Schutzschalter MTP Pumpe eines neuen Gerats befinden sich in OFF Position Schalten Sie MTP auf ON M nach Zuruck setzen von MTP Es ist ein Pumpen problem aufgetreten...

Page 62: ...eine Blockierung im Kaltetragerkreislauf befindet Verschmutzter Verdampfer Reinigen Sie den Verdampfer oder kontaktieren Sie ggf den Kunden dienst C1OP C6OP Kompressor o ldruck Alarm oder Oel stand Al...

Page 63: ...s erscheint die Bezeichnung rSt wenn der Alarm zuru ckgesetzt werden kann NO wenn dies nicht mo g lich ist Benutzen Sie die Tasten oder um durch die Alarmliste zu blattern 5 Dru cken Sie die Taste wen...

Page 64: ...000 18260 35 26102 48 20410 39 28252 53 23910 46 31752 61 131 136 145 PCGE 880 400 3 50 88000 23092 42 32516 59 26392 49 35816 66 31892 59 41316 79 138 146 158 PCGE 1000 400 3 50 103000 26490 52 38058...

Page 65: ...150 400 7 0 5 0 150 400 300 1 18300 PCGE 730 2 8 2 3 120 300 4 9 3 2 150 400 7 0 5 0 150 400 300 1 18300 PCGE 880 3 3 2 1 150 400 5 5 3 6 120 400 8 0 6 3 150 400 500 2 24400 PCGE 1000 3 5 2 1 150 400...

Page 66: ...X PCGE 590 470 R407C X X X X PCGE 730 800 R407C X X X X X PCGE 880 850 R407C X X X X X PCGE 1000 1050 R407C X X X X X PCGE 1100 1075 R407C X X X X X X PCGE 1300 1780 R407C X X X X X X PCGE 1900 1840...

Page 67: ...HAUPTSCHALTER FU R DEN KUNDEN MAIN SWITCH CUSTOMER CURE INTERRUPTOR GENERAL CUIDADO POR EL CLIENTE INTERRUTTORE GENERALE A CURA DEL CLIENTE IAF LUFTFILTER U BERWACHUNG AIRFILTER MONITORING SUPERVISION...

Page 68: ...OVERFLOW PROTECTION PROTECCION DEL DESBORDAMENTO SCARICO VASCA ANTIALLAGAMENTO TD DIGITALTHERMOMETER DIGITAL THERMOMETER TERMOMETRO DIGITAL TERMOSTATO DIGITALE TH TEMPERATURREGLER THERMOSTAT TERMOSTAT...

Page 69: ...69 11 2 WIRING DIAGRAM MODELS PCGE 11 to PCGE 25 PCGEZ 08 230V 1Ph 50Hz POWER COMMAND DIAGRAM...

Page 70: ...70 MODEL PCGE 28 230V 1Ph 50Hz POWER COMMAND DIAGRAM...

Page 71: ...71 MODEL PCGE 28 400V 3Ph 50Hz POWER COMMAND DIAGRAM...

Page 72: ...72 MODELS PCGE 31 to PCGE 41 OCGE 31 to OCGE 41 230V 1Ph 50Hz POWER SECTION...

Page 73: ...73 MODELS PCGE 31 to PCGE 41 OCGE 31 to OCGE 41 230V 1Ph 50Hz COMMAND SECTION...

Page 74: ...74 MODELS PCGE 31 to PCGE 141 OCGE 31 to OCGE 141 PCGEZ 37 to PCGEZ 100 400V 3Ph 50Hz POWER SECTION...

Page 75: ...75 MODELS PCGE 31 to PCGE 141 OCGE 31 to OCGE 141 PCGEZ 37 to PCGEZ 100 400V 3Ph 50Hz COMMAND SECTION...

Page 76: ...76 MODELS PCGE 200 to PCGE 520 OCGE 150 to OCGE 240 PCGEZ 160 to PCGEZ 300 400V 3Ph 50Hz POWER SECTION...

Page 77: ...77 MODELS PCGE 200 to PCGE 520 OCGE 150 to OCGE 240 PCGEZ 160 to PCGEZ 300 400V 3Ph 50Hz COMMAND SECTION...

Page 78: ...78 MODELS PCGE 590 to PCGE 3300 PCGEZ 360 to PCGEZ 480 400V 3Ph 50Hz...

Page 79: ...79...

Page 80: ...80...

Page 81: ...81...

Page 82: ...82 11 3 COOLING HYDRAULIC DIAGRAM MODELS PCGE 11 to PCGE 25 PCGEZ 08 with tank and pump MODEL PCGE 28 with tank and pump...

Page 83: ...83 MODELS PCGE 31 to PCGE 141 PCGEZ 37 to PCGEZ 100 with tank and pump MODELS PCGE 150 to PCGE 290 PCGEZ 160 with tank and pump...

Page 84: ...84 MODELS PCGE 370 to PCGE 520 PCGEZ 180 to PCGEZ 300 with tank and pump MODEL PCGE 590 with tank and pump...

Page 85: ...85 MODELS PCGE 730 to PCGE 880 PCGEZ 360 to PCGEZ 480 with tank and pump MODEL PCGE 1000 with tank and pump...

Page 86: ...86 MODEL PCGE 1100 with tank and pump MODEL PCGE 1300 with tank and pump...

Page 87: ...87 MODEL PCGE 1900 2200 with tank and pump MODEL PCGE 2900 3300 with tank and pump...

Page 88: ...88 MODEL PCGE 31 to 141 OCGE 31 to 141 PCGEZ 37 to PCGEZ 100 without tank MODELS PCGE 150 to PCGE 290 OCGE 150 to OCGE 240 PCGEZ 160 without tank...

Page 89: ...89 MODELS PCGE 370 to PCGE 520 PCGEZ 180 to PCGEZ 300 without tank MODEL PCGE 590 without tank...

Page 90: ...90 MODELS PCGE 730 to PCGE 880 PCGEZ 360 to PCGEZ 480 without tank MODEL PCGE 1000 without tank...

Page 91: ...91 MODEL PCGE 1100 without tank MODEL PCGE 1300 without tank...

Page 92: ...92 MODEL PCGE 1900 2200 without tank MODEL PCGE 2900 3300 without tank...

Page 93: ...etrafluorethane 743 C 1300 0 R407c 1 A1 R32 R125 134a 23 25 52 704 C 1600 0 R404a 1 A1 R125 R134a R143a 44 4 52 3750 0 REFRIGERANT HAZARDS FIRST AID PHRASE R S R134a High concentration can cause asphy...

Page 94: ...producto al cual esta declaracion se refiere es conforme a las siguientes normas y otros documentos normativos P Declaramos sob nossa exclusiva responsabilidade que o produto a que se refere esta decl...

Page 95: ...ks shall be recorded in the system booklet Verordnung EG Nr 2037 2000 des Europa ischen Parlaments und des Rates vom 29 Juni 2000 uber Stoffe die zum Abbau der Ozonschicht fuhren Der Systemmanager mus...

Page 96: ...15 INDIVIDUAL SPECIFICATION You will find the individual specification for the unit in the supplement which accompanies these operating and maintenance instructions and which forms part of this manua...

Page 97: ...97 710 1789 00 00 02B 07 05 2008...

Page 98: ...00 OCGE 31 OCGE 240 PCGEZ 08 PCGEZ 480 English Operating and maintenance manual Deutsch Bedienungs und Wartungsanleitung National Lab GmbH Grambeker Weg 157 D 23879 Moelln Phone 49 0 45 42 84 91 10 Fa...

Page 99: ...RS 17 6 2 1 OIL REQUIREMENTS 17 6 2 2 CHECKLISTE FOR OCGE 31 TO OCGE 240 17 6 3 CHECK LIST FOR COMMISSIONING GLYCOL RECIRCULATING CHILLERS 18 6 3 1 REQUIREMENTS FOR THE WATER GLYCOL MIXTURE 18 6 3 2 C...

Page 100: ...y of cooling medium e g water water glycol moisture or oil constantly within certain temperature limits operating range with the aid of a technical cooling process The most commonly used cooling mediu...

Page 101: ...in this connection The side panels of the chiller must not be removed unless this is absolutely necessary Even if the side panels are only partially removed isolate the chiller from the electricity su...

Page 102: ...f fully hermetic design with integrated bimetal switch and oil sight glass Semi hermetic compressor Semi hermetic compressor with integrated bimetal switch and oil sight glass Reciprocating piston com...

Page 103: ...gs such as non return valves filter drier sight glass and service connections 3 2 2 OPTIONS REFRIGERATION CIRCUIT Condenser air filter aluminium reusable The condenser air filter protects the condense...

Page 104: ...return circuit systems with single and dual circuit The digital thermometer shows the inlet temperature of the cooling medium VNR Non return valve in the supply line These valves are used in open syst...

Page 105: ...ite failure report summarizes all fault messages from the compressor the pump s fan s of the high pressure and low pressure pressostats the flow monitor as well as the breakage of probes Voltage free...

Page 106: ...r by means of codes 3 4 2 OPTIONS ELECTRIC CIRCUIT Connecting cable with CEE plug A ready to use mains cable is supplied with 6m oilflex cable and CEE plug 4 pole 3P PEN or 5 pole 3P N PE Compressor w...

Page 107: ...GE 81 111 760 760 1330 1250 767 100 972 100 84 1445 1 PCGEZ 100 760 760 1330 1250 767 100 972 100 84 1445 1 PCGE 141 OCGE 141 PCGEZ 160 760 760 1380 1300 767 100 972 100 84 1510 1 MODELS PCGE 150 to P...

Page 108: ...r lifting it in accordance with the applicable safety standards If the chiller is not fitted with eyebolts the lifting gear must be appropriate for the weight of the chiller as indi cated on the ident...

Page 109: ...ting it against direct sunlight You should avoid connecting the axial ventilators to air ducts as they do not have sufficient compression Higher ambient temperatures than those given as a reference va...

Page 110: ...allow use of the consumer circuit if you have to remove the chiller from the circuit switch it off or use it in different ways at the same time A pipeline system with integrated filter near the chille...

Page 111: ...atures must be observed Make sure that the mains voltage and the frequency match the data on the identification plate of the chiller and agree with those in the table in the supplement and that the pe...

Page 112: ...plate on the motor 6 1 2 CHECKLIST FOR PCGE 11 TO PCGE 3300 Controller Type A see 6 4 PCGE 11 to PCGE 28 1 Remove the lid panel and the two side panels 2 Fill the collection tank Open the tank cap and...

Page 113: ...ches MTC and MTP are in position 0 8 Fill the tank to the maximum upper arrow level indicator is located on the bottom right of the frame See Section 6 1 1 Water requirements 9 Move the motor protecti...

Page 114: ...re required by the customer 26 When the chiller is within the operating conditions set by the manufacturer check whether the current uptake of the chiller agrees with the data indicated on the identif...

Page 115: ...dinary maintenance work each chiller start should be regarded as an initial start up The initial start up calls for a number additional checks to ensure safe and reliable performance of the system 6 3...

Page 116: ...cation plate and those contained in the table in the supplement Controller Type C see 6 4 PCGEZ 240 to PCGEZ 480 1 Remove the front plate and the two side panels 2 Open the tank cap and fill the tank...

Page 117: ...hes AUX Second relay active ON A temperature alarm has been triggered ON Type B PCGE 31 to 520 OCGE 31 to 240 PCGEZ 37 to 180 Symbols degrees Celsius degrees Fahrenheit bar PSI Compressor 1 Compressor...

Page 118: ...of the chiller listed in the supplement The operating thermostat controls both the temperature adaptation functions and the other safety parameters e g alarm management and the running time of the el...

Page 119: ...the parameter you require 6 Press the key to change its value 7 Change the value with the and keys 8 Press the key again to confirm the new value After a few seconds the next parameter will be display...

Page 120: ...ressed for 10 seconds 2 Move the main switch of the chiller to position 0 Via the main switch Move the main switch to position 0 The system then switches off immediately This procedure should only be...

Page 121: ...G FOR TRANSPORT PURPOSES All work on the chiller may only be conducted by specialist personnel Follow this procedure if you need to shut down the chiller Stop the chiller and permanently disconnect it...

Page 122: ...1 Check whether the current consumption of the compressor group is within the target values indicated in these instructions 2 Clean the fin pack of the condenser using a compressed air jet with a pre...

Page 123: ...constituents must be treated in line with the provisions of the valid legislation 9 MALFUNCTIONS TROUBLESHOOTING All service and maintenance operations may only be carried out by specialized personne...

Page 124: ...tion and in case change probe HA High liquid tem perature alarm Compressor switched off by internal protection Contact a technician 2 Pump is not working fan and com pressor are working Pump is blocke...

Page 125: ...low Fill the tank A09 Pump thermal switches alarm Thermal protections of a new unit are in OFF posi tion Switch MTP in ON M after reset MTP Pump is having problem Check pump and possible obstruc tion...

Page 126: ...probe is defec tive Probe can be defective or not well connected Check connection and in case change probe A 3 Unit is work ing A13 Maintenance circuit 1 only on request Service on unit is required du...

Page 127: ...Pump thermal switches alarm Thermal protections of a new unit are in OFF posi tion Switch MTP in ON M after reset MTP Pump is having problem Reset thermal switch and in case contact of a persistent p...

Page 128: ...tches alarm Thermal protections of a new unit are in OFF po sition Switch MTC in ON M after reset MTC Compressor is having problem Reset thermal switch and in case contact of a persistent problem con...

Page 129: ...ME VON OEL RUCKKUHLERN 47 6 3 1 ANFORDERUNGEN AN DAS OEL 47 6 3 2 CHECKLISTE OCGE 31 BIS OCGE 240 47 6 4 CHECKLISTE ZUR INBETRIEBNAHME VON GLYKOL RUCKKUHLERN 48 6 4 1 ANFORDERUNGEN AN DAS WASSER GLYKO...

Page 130: ...Menge Kal tetrager z B Wasser Wasser Glykol Gemisch oder Oel mit Hilfe eines technischen Kuhlverfahrens konstant inner halb bestimmter Temperaturgrenzen Betriebsbereich zu halten Der am haufigsten ver...

Page 131: ...gung des Herstellers durchgefuhrt werden wenn Komponenten des Ge rates unter Druck stehen Verschutten Sie niemals Kaltetrager da die Gefahr von gesundheitlichen Schaden besteht Vor einer eventuel len...

Page 132: ...G Das Kuhlgerat ist bei Lieferung anschlussfertig Es besteht aus einem Gehause einem vollstandig hermetisch abge dichteten Kaltemittelkreislauf dem Kaltetrager Kreislauf und der Elektrotechnik inklusi...

Page 133: ...ialventilator zum Anschluss an einen bauseitigen Luftkanal Einspritzverfahren a Kapillareinspritzung bis zu Kuhlleistungen von ca 2 5 kW Das Kaltemittel wird uber eine definierte Kapillare eingespritz...

Page 134: ...Der Schwimmerschalter befindet sich im Tank und uberwacht den Kaltetragerstand Beim Sinken des Wasserstandes und Erreichen des Mindestgrenzwertes werden das Kuhlsystem und die Pumpe n abgeschaltet Das...

Page 135: ...Einkreis System ohne Tank ausgelegt Die Pumpe fo rdert den Kaltetrager innerhalb des Kreislaufs durch den Verdampfer und uber die Verbraucher Der Kaltetrager wird im Verdampfer abgekuhlt Der Kreislau...

Page 136: ...ltet QE Schaltschrank Im Schaltschrank sind alle elektrische Komponenten und Sicherheitsvorrichtungen enthalten TLT Betriebsthermostat Die Steuerung und Regelung der Temperatur erfolgt durch den Betri...

Page 137: ...E 81 111 760 760 1330 1250 767 100 972 100 84 1445 1 PCGEZ 100 760 760 1330 1250 767 100 972 100 84 1445 1 PCGE 141 OCGE 141 PCGEZ 160 760 760 1380 1300 767 100 972 100 84 1510 1 MODELLE PCGE 150 bis...

Page 138: ...a den geltenden Sicherheits standards zum Heben des Gerats verwendet werden Wenn das Kuhlgerat nicht mit Hebeo sen ausgestattet ist muss das Hebemittel fur das auf dem Typenschild und im Anhang angege...

Page 139: ...trachtigen empfehlen wir es vordirekter Sonneneinstrahlung zu schutzen Ein Anschluss der Axial Ventilatoren an Luftkanale sollte vermieden werden da sie keine ausreichende Pressung aufweisen Ho here R...

Page 140: ...hern zu vermeiden Die empfohlene Maschenweite fur die Kuhlgerate liegt bei 25 mit einem maximalen Druckverlust am Filter von 1 bar Die Verwendung ei nes Filters mit einer Maschenweite von 60 eignet si...

Page 141: ...t Bevor Sie das Kuhlgerat mit dem Stromnetz verbinden stellen Sie sicher dass der allgemeine Trennschalter die Anschlussleitung des Kuhlgerates vom Stromnetz trennt Fuhren Sie die Drahte durch die fle...

Page 142: ...den vom Hersteller eingestellten Betriebsbedingungen befindet ob die elektrische Absorption des Kuhlgerates mit den auf dem Typenschild angegebenen und in der Tabelle im Beiblatt enthaltenen Daten ube...

Page 143: ...ebsbedingungen erreicht hat diese sind dann erreicht wenn das Display den ein gestellten Arbeitsbereich anzeigt 17 Uberprufen Sie ob sich der angezeigte Druck innerhalb der fur die Pumpe spezifizierte...

Page 144: ...nden OCGE Serie Tank in der Nahe des Kuhlgerates max 2m Zahnradpumpe Option mit weiter entferntem Tank 2m Oel Austrittstemperatur ISO32 min 20 C ISO46 min 25 C ISO68 min 30 C ISO10 ISO68 Allgemeine Be...

Page 145: ...n Hersteller fur diesen Zweck ausdrucklich zugelassen sind Der Glykolanteil muss mindestens 20 betragen und sollte einen Wert von 25 nicht uberschreiten Ka ltetra geraustrittstemperatur Ethylenglykol...

Page 146: ...die beiden Seitenbleche 2 O ffnen Sie die den Tankdeckel und befullen Sie den Tank ggf mit einem T Stuck unter Kontrolle der Fullstandsanzeige siehe dazu Abb 1 Punkt 5 2 1 3 O ffnen Sie die Beluftungs...

Page 147: ...EIN Ein Temperaturalarm ist aufgetreten EIN Typ B PCGE 31 bis 520 OCGE 31 bis 240 PCGEZ 37 bis 180 Symbole Grad Celsius Grad Fahrenheit Bar Psi Kompressor 1 Kompressor 2 Einheit im Standby Allgemeine...

Page 148: ...eiblatt aufgefuhrten Standardspezifikationen des Kuhlgerates in vor eingestelltem Zustand geliefert Der Betriebsthermostat steuert sowohl die Temperaturanpassungsfunktionen als auch andere Sicherheits...

Page 149: ...ppe von Parametern bedeutet 2 Benutzen Sie die und Pfeile um durch die Namen der zu andernden Parametergruppen zu scrollen 3 Nachdem Sie die richtige Parametergruppe gefunden haben drucken Sie die Tas...

Page 150: ...ie den Wert mit den und Tasten 6 Drucken Sie die Taste erneut um den neuen Wert zu bestatigen Nach einigen Sekunden wird der nachste Parameter erscheinen 7 Um das Programm zu verlassen Drucken Sie die...

Page 151: ...hrt werden Wenn Sie das Kuhlgerat abschalten mussen gehen Sie wie folgt vor Stoppen Sie das Kuhlgerat und trennen Sie es dauerhaft vom Stromnetz Erst nachdem Sie es ordnungsgema vom Netz getrennt habe...

Page 152: ...eckungen der Maschinenteile Wo chentlich 1 Prufen Sie von au en ob der Stand des Kaltetragers im Tank innerhalb der Grenzen liegt die auf der An zeige an der Seite des Kuhlgerates zu sehen ist 2 Prufe...

Page 153: ...halter die Stromzufuhr zum Kuhlgerat unterbricht und bringen Sie entsprechende Hinweisschilder an um eine erneute Verbindung der Maschine zum Stromnetz zu vermeiden Zur richtigen Handhabung oder Lager...

Page 154: ...erschmutzter Verflussiger Verflussiger wie in der Wartungs anleitung beschrieben reinigen Ventilator arbeitet nicht einwandfrei oder die Dreh richtung stimmt nicht Uberprufen Sie die Sicherung des Ven...

Page 155: ...en Sie ggf den Kundendienst Luftfilter verschmutzt nur bei Geraten die uber diese Option verfugen Demontieren Sie den Luftfilter vom Gerat und reinigen Sie ihn mit Druckluft oder Wasser LA Zu niedrige...

Page 156: ...ms erset zen Sie die Steuereinheit M 2 Ventilator und Kompressor funkti onieren nicht Pumpe funktioniert A01 Hochdruck Alarm beim Kaltemittel Zu hohe Umgebungs temperatur Umgebungstemperatur senken Be...

Page 157: ...Fuhler ist defekt Fuhler moglicherweise defekt oder nicht richtig angeschlossen Anschluss uberprufen und ggf Fuhler ersetzen A 3 Gerat funktioniert A13 Wartung von Kreislauf 1 nur bei Bedarf Wartung d...

Page 158: ...schal ter Pumpe Alarm der Pumpe Der Schutzschalter MTP Pumpe eines neuen Gerats befinden sich in OFF Position Schalten Sie MTP auf ON M nach Zuruck setzen von MTP Es ist ein Pumpen problem aufgetreten...

Page 159: ...eine Blockierung im Kaltetragerkreislauf befindet Verschmutzter Verdampfer Reinigen Sie den Verdampfer oder kontaktieren Sie ggf den Kunden dienst C1OP C6OP Kompressor o ldruck Alarm oder Oel stand Al...

Page 160: ...s erscheint die Bezeichnung rSt wenn der Alarm zuru ckgesetzt werden kann NO wenn dies nicht mo g lich ist Benutzen Sie die Tasten oder um durch die Alarmliste zu blattern 5 Dru cken Sie die Taste wen...

Page 161: ...000 18260 35 26102 48 20410 39 28252 53 23910 46 31752 61 131 136 145 PCGE 880 400 3 50 88000 23092 42 32516 59 26392 49 35816 66 31892 59 41316 79 138 146 158 PCGE 1000 400 3 50 103000 26490 52 38058...

Page 162: ...150 400 7 0 5 0 150 400 300 1 18300 PCGE 730 2 8 2 3 120 300 4 9 3 2 150 400 7 0 5 0 150 400 300 1 18300 PCGE 880 3 3 2 1 150 400 5 5 3 6 120 400 8 0 6 3 150 400 500 2 24400 PCGE 1000 3 5 2 1 150 400...

Page 163: ...X PCGE 590 470 R407C X X X X PCGE 730 800 R407C X X X X X PCGE 880 850 R407C X X X X X PCGE 1000 1050 R407C X X X X X PCGE 1100 1075 R407C X X X X X X PCGE 1300 1780 R407C X X X X X X PCGE 1900 1840...

Page 164: ...HAUPTSCHALTER FU R DEN KUNDEN MAIN SWITCH CUSTOMER CURE INTERRUPTOR GENERAL CUIDADO POR EL CLIENTE INTERRUTTORE GENERALE A CURA DEL CLIENTE IAF LUFTFILTER U BERWACHUNG AIRFILTER MONITORING SUPERVISION...

Page 165: ...OVERFLOW PROTECTION PROTECCION DEL DESBORDAMENTO SCARICO VASCA ANTIALLAGAMENTO TD DIGITALTHERMOMETER DIGITAL THERMOMETER TERMOMETRO DIGITAL TERMOSTATO DIGITALE TH TEMPERATURREGLER THERMOSTAT TERMOSTAT...

Page 166: ...69 11 2 WIRING DIAGRAM MODELS PCGE 11 to PCGE 25 PCGEZ 08 230V 1Ph 50Hz POWER COMMAND DIAGRAM...

Page 167: ...70 MODEL PCGE 28 230V 1Ph 50Hz POWER COMMAND DIAGRAM...

Page 168: ...71 MODEL PCGE 28 400V 3Ph 50Hz POWER COMMAND DIAGRAM...

Page 169: ...72 MODELS PCGE 31 to PCGE 41 OCGE 31 to OCGE 41 230V 1Ph 50Hz POWER SECTION...

Page 170: ...73 MODELS PCGE 31 to PCGE 41 OCGE 31 to OCGE 41 230V 1Ph 50Hz COMMAND SECTION...

Page 171: ...74 MODELS PCGE 31 to PCGE 141 OCGE 31 to OCGE 141 PCGEZ 37 to PCGEZ 100 400V 3Ph 50Hz POWER SECTION...

Page 172: ...75 MODELS PCGE 31 to PCGE 141 OCGE 31 to OCGE 141 PCGEZ 37 to PCGEZ 100 400V 3Ph 50Hz COMMAND SECTION...

Page 173: ...76 MODELS PCGE 200 to PCGE 520 OCGE 150 to OCGE 240 PCGEZ 160 to PCGEZ 300 400V 3Ph 50Hz POWER SECTION...

Page 174: ...77 MODELS PCGE 200 to PCGE 520 OCGE 150 to OCGE 240 PCGEZ 160 to PCGEZ 300 400V 3Ph 50Hz COMMAND SECTION...

Page 175: ...78 MODELS PCGE 590 to PCGE 3300 PCGEZ 360 to PCGEZ 480 400V 3Ph 50Hz...

Page 176: ...79...

Page 177: ...80...

Page 178: ...81...

Page 179: ...82 11 3 COOLING HYDRAULIC DIAGRAM MODELS PCGE 11 to PCGE 25 PCGEZ 08 with tank and pump MODEL PCGE 28 with tank and pump...

Page 180: ...83 MODELS PCGE 31 to PCGE 141 PCGEZ 37 to PCGEZ 100 with tank and pump MODELS PCGE 150 to PCGE 290 PCGEZ 160 with tank and pump...

Page 181: ...84 MODELS PCGE 370 to PCGE 520 PCGEZ 180 to PCGEZ 300 with tank and pump MODEL PCGE 590 with tank and pump...

Page 182: ...85 MODELS PCGE 730 to PCGE 880 PCGEZ 360 to PCGEZ 480 with tank and pump MODEL PCGE 1000 with tank and pump...

Page 183: ...86 MODEL PCGE 1100 with tank and pump MODEL PCGE 1300 with tank and pump...

Page 184: ...87 MODEL PCGE 1900 2200 with tank and pump MODEL PCGE 2900 3300 with tank and pump...

Page 185: ...88 MODEL PCGE 31 to 141 OCGE 31 to 141 PCGEZ 37 to PCGEZ 100 without tank MODELS PCGE 150 to PCGE 290 OCGE 150 to OCGE 240 PCGEZ 160 without tank...

Page 186: ...89 MODELS PCGE 370 to PCGE 520 PCGEZ 180 to PCGEZ 300 without tank MODEL PCGE 590 without tank...

Page 187: ...90 MODELS PCGE 730 to PCGE 880 PCGEZ 360 to PCGEZ 480 without tank MODEL PCGE 1000 without tank...

Page 188: ...91 MODEL PCGE 1100 without tank MODEL PCGE 1300 without tank...

Page 189: ...92 MODEL PCGE 1900 2200 without tank MODEL PCGE 2900 3300 without tank...

Page 190: ...etrafluorethane 743 C 1300 0 R407c 1 A1 R32 R125 134a 23 25 52 704 C 1600 0 R404a 1 A1 R125 R134a R143a 44 4 52 3750 0 REFRIGERANT HAZARDS FIRST AID PHRASE R S R134a High concentration can cause asphy...

Page 191: ...producto al cual esta declaracion se refiere es conforme a las siguientes normas y otros documentos normativos P Declaramos sob nossa exclusiva responsabilidade que o produto a que se refere esta decl...

Page 192: ...ks shall be recorded in the system booklet Verordnung EG Nr 2037 2000 des Europa ischen Parlaments und des Rates vom 29 Juni 2000 uber Stoffe die zum Abbau der Ozonschicht fuhren Der Systemmanager mus...

Page 193: ...15 INDIVIDUAL SPECIFICATION You will find the individual specification for the unit in the supplement which accompanies these operating and maintenance instructions and which forms part of this manua...

Page 194: ...97 710 1789 00 00 02B 07 05 2008...

Reviews: