NI USB-443x Calibration Procedure

Test Equipment

Table lists the equipment National Instruments recommends for calibrating the NI USB-443

x

.

If the recommended equipment is not available, select a substitute calibration standard that meets the 
given specifications.

Table 1.  

Recommended Equipment

Equipment

Recommended Model

Parameter 

Measured/ 

Adjusted

Minimum Requirements

Calibrator

Fluke 5700A

AI AC Coupled 
Gain

AI DC Coupled 
Gain

AI Offset

Frequency Range: 1 kHz

Voltage Range: up to 9 V

pk

AC Voltage Accuracy: ± 0.03% at 
1 kHz

*

DC Voltage Accuracy: 
±150 ppm at 9 V 
±100

μ

V with 1 mA load current

Function Generator

HP/Agilent 33250A

Timebase 
Frequency 
Accuracy

Frequency Range: up to 10 kHz

Frequency Accuracy: ±5 ppm

Voltage Range: up to 9 V

pk

Digital Multimeter

NI 4071

AO Gain

AO Offset

DC Voltage Resolution: 1

μ

DC Voltage Accuracy at 3.15 V: 
±150 ppm

BNC Shorting Cap

NI USB-4431: 
quantity 4

NI USB-4432: 
quantity 5

Pomona Electronics 5085

AI Offset

Resistance: <1

Ω

BNC Cables: 
quantity 6

Pomona Electronics 5697

All

Characteristic Impedance: 50

Ω

BNC (Female) to 
Banana Adapter

Pomona Electronics 1269

All

BNC T-Connectors: 
quantity 4

Pomona Electronics 4896

AI Gain

AI Offset

Characteristic Impedance: 50

Ω

*

 AC accuracy is the sum of all errors, including percent of reading error and temperature error, but excluding percent of 

range error.

 Frequency accuracy is the sum of all errors, including initial accuracy and stability errors.

Summary of Contents for USB-443 Series

Page 1: ...tion Procedures 4 Initial Setup 4 Changing Coupling on NI USB 443x Devices 5 Verification 6 Analog Input Performance Verification 6 AI Offset Verification 6 AI AC Coupled Gain Accuracy Verification 7...

Page 2: ...ed guides provides instructions for installing and configuring NI USB 443x devices NI DAQmx C Reference Help contains the C reference and general information about measurement concepts NI USB 443x Spe...

Page 3: ...Function Generator HP Agilent 33250A Timebase Frequency Accuracy Frequency Range up to 10 kHz Frequency Accuracy 5 ppm Voltage Range up to 9 Vpk Digital Multimeter NI 4071 AO Gain AO Offset DC Voltag...

Page 4: ...e device and configure it in Measurement Automation Explorer MAX 2 Verification Verify the existing operation of the device This step confirms whether the device is operating within the published spec...

Page 5: ...re is no 30 seconds wait between setting the coupling on the device and starting the acquisition For flexibility NI DAQmx gives you the ability to download channel configuration settings to hardware a...

Page 6: ...f NI USB 443x devices Refer to the NI USB 443x Specifications for the number of channels and performance specifications AI Offset Verification Note Skip the AC coupled portion of this procedure if bot...

Page 7: ...lication 7 Repeat steps 1 through 6 In step 2 set the coupling parameter to AC Note Skip the AC coupled portion of this procedure if both of the following apply Your device is hardware revision C or e...

Page 8: ...ifying only the channels used in your application AI DC Coupled Gain Accuracy Verification Complete the following steps to verify DC coupled gain accuracy 1 Connect the output of the calibrator to the...

Page 9: ...1 through 5 In step 2 output 9 VDC instead of 9 VDC Subtract the averaged value measured for 0 VDC from the 9 VDC measurement Table 7 AI DC Coupled Gain Accuracy Verification Configuration for NI USB...

Page 10: ...fy the AO offset 1 Connect channel AO 0 on the device to the input of the multimeter 2 Create and configure an AO voltage task on the NI USB 4431 as shown in Table 10 3 Create an array of length 96 00...

Page 11: ...nfigured to generate samples continuously 5 Wait 100 ms This compensates for the filter delay of the analog output channel 6 Measure the DC voltage of the analog output channel using the multimeter in...

Page 12: ...evice 2 Configure the function generator to generate a 10 kHz sine wave with an amplitude of 9 Vpk and no DC offset 3 Create and configure an AI voltage task on the NI USB 443x as shown in Table 14 4...

Page 13: ...t the output of the calibrator to all input channels on the device Use the BNC T connectors to route the signal to all of the inputs 2 Configure the calibrator to output a 1 kHz signal with a 6 3 Vrms...

Page 14: ...for NI USB 443x Configuration Value Minimum value 10 0 Maximum value 10 0 Terminal configuration Pseudodifferential Acquisition mode Continuous Rate 1000 0 Coupling AC Physical channel Dev_name aix D...

Page 15: ...cel if there has been any error during the AI calibration or if you do not want to save the new AI calibration constants in the device EEPROM Use the action commit to save the new AI calibration const...

Page 16: ...tinuously 6 Wait 100 ms This compensates for the filter delay of the analog output channel 7 Measure the DC voltage of channel AO 0 on the device using the multimeter Record this value as actualHighVo...

Page 17: ...device EEPROM The analog output of the device is now calibrated with respect to your external equipment After calibrating the device verify the analog output operation by repeating the Analog Output P...

Page 18: ...rt export data 2011 National Instruments Corporation All rights reserved 373489A 01 Oct11 Where to Go for Support The National Instruments Web site is your complete resource for technical support At n...

Reviews: