National Instruments Corporation
17
NI USB-443x Calibration Procedure
10. Perform AO calibration using the DAQmx Adjust DSA AO Calibration VI as shown below.
11. Finish the AO calibration with the DAQmx Close External Calibration VI as shown below. Use the
action
cancel
if there has been any error during the AO calibration or if you do not want to save
the new AO calibration constants in the device EEPROM. Use the action
commit
to save the new
AO calibration constants in the device EEPROM.
The analog output of the device is now calibrated with respect to your external equipment. After
calibrating the device, verify the analog output operation by repeating the
section.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxAdjustDSAAOCal
with the following
parameters:
calhandle
:
myCalHandle
channel
:
0
requestedHighVoltage
: 2.0
actualHighVoltage
:
(value from step 7)
requestedLowVoltage
: –2.0
actualLowVoltage
:
(value from step 9)
gainSetting
: 0.0
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCloseExtCal
with the following
parameters:
calHandle
:
myCalHandle
action
:
DAQmx_Val_Action_Commit
or
DAQmx_Val_Action_Cancel