NI PXIe-5693 Calibration Procedure
|
© National Instruments
|
31
•
Detector type: Auto Peak
•
Average count: 50
•
External reference source: EXT
13. Configure the RF source 1 and RF source 2 according to the following settings:
•
Single frequency mode: Enabled
•
Power level: -5 dBm
14. Set the power sensor A settling percentage to 0.1% and program the power meter to trigger
when settled.
15. Set the NI 5693 center frequency according to the first row in Table 10. This is the
IMD Frequency
.
Table 10.
Out-of-Band TOI Verification Test Frequencies
IMD Frequency (MHz)
Tone Spacing (MHz)
IMD Side
30
20
Lower
50
40
Lower
90
40
Lower
110
40
Upper
150
40
Lower
170
50
Upper
190
50
Lower
240
60
Upper
340
60
Lower
360
70
Upper
550
70
Lower
590
80
Upper
930
80
Lower
960
200
Upper
1,540
410
Lower
1,580
120
Upper
1,980
130
Lower
2,020
140
Upper
2,490
150
Lower