NI PXIe-5693 Calibration Procedure
Figure 6.
Spectrum Analyzer Noise Figure Characterization Equipment Setup
3.
Create a new session for the NI 5693.
4.
Configure the NI 5693 according to the following settings:
•
Reference level: -60 dBm
•
Preamp enabled: Enabled
•
IF output power level: -10 dBm
5.
Configure the spectrum analyzer according to the following settings:
•
Span: 0 Hz
•
Reference level: -50 dBm
•
Resolution bandwidth filter: 10 kHz
•
Video bandwidth filter: 30 Hz
•
RF attenuation: 0 dB
•
Preamp: On
•
Trace type: Average
•
Detector type: Sample
•
Average count: 100
6.
Turn off the spectrum analyzer noise source control.