36
|
ni.com
|
NI PXIe-5693 Calibration Procedure
•
Average count: 5
•
External reference source: EXT
10. Configure the RF source 1 according to the following settings:
•
Single frequency mode: Enabled
•
Power level: 0 dBm
11. Set the power sensor A settling percentage to 0.1% and program the power meter to trigger
when settled.
12. Set the NI 5693 center frequency to the center frequency in the first row of Table 13.
13. Enable the NI 5693 filter 1 path.
14. Commit the NI 5693 settings to hardware.
15. Set the spectrum analyzer frequency to the source frequency in the first row Table 13.
16. Set the RF Source 1 center frequency to the source frequency in the first row of Table 13.
Table 13.
Reverse Isolation Test Frequencies
Filter
Center Frequency (MHz)
Source Frequency (MHz)
1
27
4,639.5
2
47
4,659.5
3
80
4,692.5
4
130
4,742.5
5
193
4,805.5
6
288
4,900.5
7
463
5,075.5
8
763
5,375.5
9
1,255
5,867.5
10
1,780
6,392.5
11
2,250
6,862.5
12
2,750
7,362.5
13
3,400
4,012.5
14
4,200
4,812.5
15
5,200
5,812.5
16
6,400
7,012.5