Figure 2. Reference Amplitude Accuracy Verification (10 MHz) Equipment Setup
250 kHz–20 GHz Signal Generator
ACCESS
ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50
Ω
1
2
3
1. PXIe-5654 RF Analog Signal Generator
2. Power Meter
3. Power Sensor
2.
Create a new device session for the PXIe-5654.
3.
Configure the PXIe-5654 using the following settings:
•
Frequency: 4 GHz
•
Power Level: 0 dBm
•
Reference Clock Export Output Terminal: RefOut
4.
Initiate signal generation.
5.
Disable the RF output.
6.
Check the signal generation status and verify that there are no reported errors or
warnings.
7.
Measure the PXIe-5654 REF OUT 10 MHz front panel connector power using the power
meter.
8.
Correct the power meter reading for the 10 MHz measurement frequency and ensure that
the corrected power is between 3 dBm and 7 dBm.
9.
Close the device session.
PXIe-5654 Calibration Procedure
|
© National Instruments
|
11