Verifying RF OUT Frequency Accuracy
This procedure verifies that the internal signal generation circuitry is adjusted for correct
frequency accuracy.
1.
Connect the PXIe-5654 RF OUT front panel connector to the signal source analyzer RF
input connector using the K(m)-to-K(m) cable as shown in the following figure.
Figure 5. RF OUT Frequency Accuracy Verification Equipment Setup (Reference
Connections Not Shown)
250 kHz–20 GHz Signal Generator
ACCESS
ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50
Ω
1
2
3
1. PXIe-5654 RF Analog Signal Generator
2. Signal Source Analyzer
3. K(m)-to-K(m) Cable
2.
Create a new device session for the PXIe-5654.
3.
Create a list of test frequencies, including endpoints, containing 100 MHz and 1 GHz,
according to the frequency range of your PXIe-5654 model, with 1 GHz increments.
PXIe-5654 Calibration Procedure
|
© National Instruments
|
15