Generating a Signal Using the NI-RFSG Soft Front
Panel

To verify your device configuration, use the NI-RFSG Soft Front Panel (SFP) in MAX to
generate a simple signal.
1.

Within MAX, select the PXIe-5654 RF signal generator module in the configuration tree.

2.

Select 

Soft Front Panel

 from the MAX toolbar.

The NI-RFSG SFP launches.

3.

Within the NI-RFSG SFP, specify a frequency and a power level for signal generation.

Caution

 

Clicking 

RF On/Off

 generates a signal from the RF OUT connector

of the PXIe-5654 front panel. Disconnect any equipment that can be damaged
by the test signal prior to clicking the 

RF On/Off

 button on the NI-RFSG SFP.

4.

Click 

RF On/Off

 to begin signal generation.

Note

 

Refer to the 

Troubleshooting

 section of this document if an

ACTIVE LED does not turn on or if the NI-RFSG SFP generates an error.

During signal generation, the ACTIVE LED on the PXIe-5654 hardware module
illuminates.

5.

Click 

RF On/Off

 to stop signal generation.

Building a Basic NI-RFSG Application

You can build a basic NI-RFSG application in LabVIEW for generating continuous sine wave
signals.

Adding the Core NI-RFSG VIs to a Blank VI

Create a new application that includes the core NI-RFSG VIs.
1.

Launch LabVIEW.

2.

To create a blank VI, select 

File

»

New VI

.

3.

Display the block diagram by selecting 

Window

»

Show Block Diagram

.

Tip

 

Activate the LabVIEW context help by selecting 

Help

»

Show Context

Help

.

4.

Right-click the VI block diagram to launch the Functions palette.

5.

Navigate to the NI-RFSG VIs on the 

NI-RFSG

 palette.

Tip

 

You can use the 

Search

 button on the 

Functions

 palette to find the 

NI-

RFSG

 palette.

6.

Add the core NI-RFSG VIs from the NI-RFSG palette to the block diagram, and wire the
VIs together as shown in the following figure.

12

  | 

ni.com

  | 

PXIe-5654 Getting Started Guide

Summary of Contents for PXIe-5654

Page 1: ...ons et respectez toutes les mises en garde de la documentation d utilisation L utilisation du produit de toute autre fa on que celle sp cifi e risque de l endommager et de compromettre la protection d...

Page 2: ...nuals for more information about minimum system requirements recommended system and supported application development environments ADEs Unpacking the Kit Notice To prevent electrostatic discharge ESD...

Page 3: ...ce kit that you need to install or operate the PXIe 5654 Required Items A PXI Express chassis and chassis documentation The PXIe 1085 chassis is one available option for your device For more informati...

Page 4: ...about installing removing and upgrading NI software using NI Package Manager 3 Follow the instructions in the installation prompts to install the NI RFSG driver software Note Windows users may see acc...

Page 5: ...Chassis Compatibility Symbols 1 2 3 4 5 1 PXI Express System Controller Slot 2 PXI Peripheral Slot 3 PXI Express Hybrid Peripheral Slot 4 PXI Express System Timing Slot 5 PXI Express Peripheral Slot P...

Page 6: ...els depending on your application Note For more information about installing slot blockers and filler panels go to ni com r pxiblocker 12 Power on the chassis Related Information Installing the Softwa...

Page 7: ...ded to complete the tightening of the SMA cable to 1 N m Rotate the torque wrench no more than one revolution to tighten the SMA cable before reaching maximum torque Related Information Refer to the P...

Page 8: ...2 dBm NOM REF OUT 2 100 MHz 5 dBm 2 dBm NOM TRIG IN OUT 3 3V CMOS IN OUT 0 5V MIN IN 5 5V MAX IN ESD SENSITIVE ALL PORTS 50 Table 1 Device Front Panel Icon Definitions Refer to the user documentation...

Page 9: ...at accepts a modulating square wave to pulse the RF output FM IN Input terminal for frequency modulation AM IN Input terminal for amplitude modulation ALC IN Input terminal that connects to the ALC OU...

Page 10: ...pand your Chassis tree item MAX lists all modules installed in the chassis Your default names may vary Note If you do not see your module listed press F5 to refresh the list of installed modules If th...

Page 11: ...ation in Microsoft Visual C C Related Information Refer to the Getting Started topic of the NI RF Signal Generators Help for instructions about how to generate signals in a specific ADE Refer to the C...

Page 12: ...NI RFSG SFP generates an error During signal generation the ACTIVE LED on the PXIe 5654 hardware module illuminates 5 Click RF On Off to stop signal generation Building a Basic NI RFSG Application Yo...

Page 13: ...he VI front panel power level dBm control enter 0 In the frequency Hz control enter 100M 100 MHz 12 In the VI front panel resource name control enter the PXIe 5654 device name that you specified in MA...

Page 14: ...output of the status element to an input of the Or function 9 Right click the unused input of the Or function and select Create Control to create a Boolean control 10 In the VI front panel right clic...

Page 15: ...specified in MAX in the resource name control 4 Click the Run button on the toolbar to initiate sine wave generation 5 Click the VI front panel STOP button to stop sine wave generation You have succes...

Page 16: ...it for damage Do not reinstall a damaged device 4 Reinstall the module in a different chassis slot 5 Power on the chassis 6 Verify that the device appears in MAX 7 Reset the device in MAX and perform...

Page 17: ...downloads and troubleshooting and application development self help such as tutorials and examples Visit ni com services to learn about NI service offerings such as calibration options repair and rep...

Page 18: ...LAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how...

Reviews: