Table 1. PXIe-5164 Test Equipment
Equipment
Recommended
Model
Where Used
Minimum Requirements
Oscilloscope
calibrator
Fluke 9500B/600
with Fluke 9530
Active Head
Verifications:
•
Timebase
accuracy
•
DC accuracy
•
Input
capacitance
Adjustment
Sine wave generation:
•
Amplitude: 0.9 V
pk-pk
into
50 Ω
•
Frequency: 11 MHz and
99 MHz
•
Frequency accuracy:
±0.25 ppm
Square wave generation:
•
Amplitude: 0.5 V
pk-pk
to
45 Vpk-pk into 1 MΩ
symmetrical to ground (0 V)
•
Frequency: 500 Hz
•
Abberations: <2% of peak
for the first 500 ns
DC generation:
•
Amplitude: ±2.5 V into
50 Ω, ±200 V into 1 MΩ
•
Accuracy:
±(0.025% of
25 µV)
Capacitance measurement:
•
Range: 15 pF to 25 pF
•
Accuracy: ±(2% of reading
± 0.25 pF)
PXIe-5164 Calibration Procedure
|
© National Instruments
|
3