Table 1. PXIe-5164 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
BNC (m)-to-
BNC (m)
cable (x2)
Pasternack PE308
Verifications:
•
AC amplitude
accuracy
Length: ≤1 meter
Power sensor Rohde & Schwarz
NRP-Z91
Test system
characterization
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
Power measurement:
•
Frequency range: 50 kHz to
401 MHz
•
Power range: -16 dBm to
10 dBm
•
VSWR: <1.11
•
Absolute accuracy
–
<0.048 dB for 50 kHz
to <100 MHz
–
<0.063 dB for
100 MHz to 401 MHz
•
Relative accuracy at -4 dBm
–
<0.022 dB for 50 kHz
to <100 MHz
–
<0.022 dB for
100 MHz to 401 MHz
Signal
generator
Rohde & Schwarz
SMA100A
Test system
characterization
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
Sine wave generation:
•
Amplitude: -10 dBm to
16 dBm
•
Frequency: 50 kHz to
401 MHz
•
Harmonics: <-30 dBc
•
Frequency accuracy:
±100.0 ppm
PXIe-5164 Calibration Procedure
|
© National Instruments
|
5