Reverification
Repeat the
section to determine the as-left status of the PXIe-5164.
Note
If any test fails reverification after performing an adjustment, verify that you
have met the test conditions before returning your PXIe-5164 to NI. Refer to the
Worldwide Support and Services
section for information about support resources or
service requests.
Updating Verification Date and Time
This procedure updates the date and time of the last PXIe-5164 verification.
Prior to updating the verification date and time, you must successfully complete all required
verifications or reverifications following adjustment.
Call the niHSAI Calibration API v1 Host.lvlib:Set Verification Date and Time VI with the
following settings:
•
Wire the current date and time to the
verification date
parameter.
•
Wire the current calibration password to the
calibration password
parameter. The
default password is
NI
.
Worldwide Support and Services
The NI website is your complete resource for technical support. At
access to everything from troubleshooting and application development self-help resources to
email and phone assistance from NI Application Engineers.
Visit
for NI Factory Installation Services, repairs, extended warranty, and
other services.
Visit
to register your NI product. Product registration facilitates technical
support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the
European Communities using the manufacturer’s declaration of conformity. This system
affords the user protection for electromagnetic compatibility (EMC) and product safety. You
can obtain the DoC for your product by visiting
calibration, you can obtain the calibration certificate for your product at
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas,
78759-3504. NI also has offices located around the world. For telephone support in the United
States, create your service request at
or dial 1 866 ASK MYNI (275 6964). For
telephone support outside the United States, visit the
Worldwide Offices
section of
to access the branch office websites, which provide up-to-date contact information,
support phone numbers, email addresses, and current events.
PXIe-5164 Calibration Procedure
|
© National Instruments
|
51