Shock and Vibration
Operating shock .....................................................30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC-60068-2-27.
Meets MIL-PRF-28800F Class 2 limits.)
Random vibration
Operating .......................................................5 Hz to 500 Hz, 0.3 g
rms
Nonoperating .................................................5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC-60068-2-64.
Nonoperating test profile exceeds the requirements
of MIL-PRF-28800F, Class 3.)
Safety Voltages
Refer to the
NI PXIe-4353 Specifications
document for maximum voltage limits.
Isolation
Channel-to-channel ........................................None
Channel-to-earth ground ................................300 V
rms
, Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical
distribution system. This category refers to local-level electrical distribution, such as that provided by a
standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Caution
Do
not
connect the TB-4353 to signals or use for measurements within Measurement
Categories III or IV.
Caution
The protection provided by the TB-4353 can be impaired if it is used in a manner not
described in this document.
Safety Standards
This product meets the requirements of the following standards of safety for electrical equipment for
measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for
measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note
In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial,
light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per
CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.