NI PXI-4130 Calibration Procedure
|
© National Instruments
|
9
Table 5.
NI 4130 Output Parameters and Test Limits for Voltage Programming
Accuracy Verification
Channel
Voltage Level
Range (V)
Iteration
Output (V)
Test Limit (mV)
0
6
1
0
±4.00
2
1.5
±4.75
3
3
±5.50
4
4.5
±6.25
5
6
±7.00
1
6
1
0
±1.50
2
1.5
±2.01
3
3
±2.52
4
4.5
±3.03
5
6
±3.54
6
-1.5
±2.01
7
-3
±2.52
8
-4.5
±3.03
9
-6
±3.54
20
1
0
±1.80
2
5
±3.50
3
10
±5.20
4
15
±6.90
5
20
±8.60
6
-5
±3.50
7
-10
±5.20
8
-15
±6.90
9
-20
±8.60