NI PXI-4130 Calibration Procedure
|
© National Instruments
|
7
Verification of the NI 4130 is complete only after you successfully complete all tests in this
section.
Note
If verification fails post-adjustment, confirm that you have met the required
before you return the NI 4130 to NI
for repair.
Table 4.
Calibration Equipment Configuration for Current Programming and
Measurement Verification/Adjustment
NI 4130
DMM
*
Calibrator
Resistance
Channel
Range
Function
Range
Input
Impedance
†
Resolution
in Digits
0
1 A
DC
Current
1 A
N/A
6.5
N/A
1
200
μ
A
DC
Voltage
10 V
10 G
Ω
7.5
10 k
Ω
2 mA
DC
Voltage
10 V
10 G
Ω
7.5
1 k
Ω
20 mA
DC
Voltage
10 V
10 G
Ω
7.5
100
Ω
200 mA
DC
Voltage
10 V
10 G
Ω
7.5
10
Ω
2 A
DC
Current
3 A
N/A
6.5
N/A
* Use the highest resolution available on the DMM.
†
Assumes an NI 4071 DMM. For all other DMMs, use the range and input impedance closest to the
values listed in this table.