![background image](http://html1.mh-extra.com/html/national-instruments/pxi-4130/pxi-4130_calibration-procedure_3594873043.webp)
NI PXI-4130 Calibration Procedure
|
© National Instruments
|
43
Table 21.
NI 4130 Output Parameters and Test Limits for Current Measurement
Accuracy Adjustment
Channel
Current
Limit
Range
Voltage
Level
Range
Voltage
Level
Iteration
Output
0
1 A
6 V
0 V
1
350 mA
6 V
2
350 mA
3
700 mA
1
200
μ
A
20 V
0 V
1
200
µ
A
20 V
2
-20 V
3
2 mA
0 V
1
2
m
A
20 V
2
-20 V
3
20 mA
0 V
1
20 mA
20 V
2
-20 V
3
200 mA
0 V
1
200 mA
20 V
2
-20 V
3
2 A
0 V
1
800 mA
20 V
2
-20 V
3