Your device includes a 16 position rotary switch for setting its IP address. The factory default
address for each switch setting is shown in the table below.
The IP address corresponding to positions 1 through E can be changed using the NI Stepper
Configuration Utility software. Setting 0 is always 10.10.10.10, the universal recovery address.
Setting F is DHCP, which commands the device to get an IP address from a DHCP server on the
network. The IP address automatically assigned by the DHCP server may be dynamic or static
depending on how the administrator has configured DHCP. The DHCP setting is reserved for
advanced users.
Your PC, or any other equipment that you use to communicate with the device, will also have
a unique address.
On the device switch settings 1 through E use the standard class B subnet mask (i.e., 255.255.0.0).
The mask for the universal recovery address is the standard class A (i.e., 255.0.0.0).
Table 2.
IP Address Rotary Switch Settings
Position
IP Address
0
10.10.10.10
1
192.168.1.10
2
192.168.1.20
3
192.168.1.30
4
192.168.0.40
5
192.168.0.50
6
192.168.0.60
7
192.168.0.70
8
192.168.0.80
9
192.168.0.90
A
192.168.0.100
B
192.168.0.110
C
192.168.0.120
D
192.168.0.130
E
192.168.0.140
F
DHCP