Storage Environment
Ambient temperature range ..............................-20 to 70 °C
(Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2.)
Relative humidity range....................................5% to 95% noncondensing
(Tested in accordance with IEC 60068-2-56.)
Shock and Vibration
Operational Shock.............................................30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC 60068-2-27.
Test profile developed in accordance with
MIL-PRF-28800F.)
Random Vibration
Operating ..................................................5 to 500 Hz, 0.3 g
rms
Nonoperating ............................................5 to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC 60068-2-64.
Nonoperating test profile exceeds the
requirements of MIL-PRF-28800F, Class 3.)
Caution
Clean the NI PXI-665
x
module with a soft nonmetallic brush. Make sure
that the device is completely dry and free from contaminants before returning it to
service.
Safety
This product is designed to meet the requirements of the following standards of safety for
electrical equipment for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.