background image

©

 National Instruments Corporation

7

Getting Started with the NI PCIe-1429 and NI PCIe-1430

5.

Right-click 

Port 0 

or 

Port 1 

and select 

Camera

.

6.

Select your camera from the menu. If your camera is not in the menu, verify that the appropriate 
camera file is installed in the 

<NI-IMAQ>\Data

 directory.

Note

If you lose communication with the camera during an acquisition for any reason, such as 

unplugging a cable or powering off your camera, you must restart the acquisition to allow the device 
to relock the incoming timing signals.

Note

The NI 1429 and NI 1430 are each equipped with a temperature monitor. The device will stop 

acquiring images if it reaches its maximum allowable temperature. An error will occur, reporting that 
the maximum temperature has been reached.

Specifications

The following specifications apply to the NI 1429 and NI 1430 image acquisition devices. These 
specifications are typical at 25 °C, unless otherwise stated.

External Connections

Number of external trigger I/O lines......................1

Note

External TTL lines on the 15-pin D-SUB connector become unavailable when the trigger 

source is set as an optically isolated input or an RS-422 input in MAX.

External trigger lines

Voltage range .................................................0 V to 5 V (TTL)

V

IH

 min ..................................................2.0 V

V

IL

 max ..................................................0.8 V

V

OH

 min .................................................2.4 V at 8 mA source

V

OL

 max .................................................0.55 V at 8 mA sink

Polarity...........................................................Programmable, active high or active low

Power-on state........................................................Input (high-impedance) 22.1 k

Ω

 pull-down to 

ground

Camera interface ....................................................Camera Link 1.1

Clocks

Pixel clock frequency range...................................20 MHz to 85 MHz

1

Note

The Camera Link specification requires cameras to transmit at a minimum of 20 MHz.

PCI Express Interface

PCI Express compliance ........................................Version 1.0a

Native link width....................................................x4

Up-plugging link width availability.......................x8, x16

Note

Some system devices limit data transfer rates for plug-in devices in an up-plugging 

configuration. Refer to the documentation provided by your computer manufacturer to determine
if your computer will support a x4 plug-in device at a x4 data rate in a larger slot.

1   

This value corresponds to the serialized Camera Link cable transmission rate of 140 to 595 MHz.

Summary of Contents for NI PCIe-1429

Page 1: ...29 Camera Link compatible camera One MDR 26 pin Camera Link cable for Base configuration cameras Two MDR 26 pin Camera Link cables of the same length for Medium or Full configuration cameras NI 1430 U...

Page 2: ...you must follow when installing and operating the device Do not operate the device in a manner not specified in the documentation Misuse of the device may result in a hazard and may compromise the sa...

Page 3: ...asurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS4 voltage This category is for measurements of voltages...

Page 4: ...rotect yourself and the computer from electrical hazards the computer must remain unplugged until the installation is complete 3 Remove the computer cover to expose the expansion slots Caution Install...

Page 5: ...il of the computer Note If you will be using the NI Camera Link I O Extension Board refer to the NI Camera Link I O Extension Board User Guide for installation instructions 8 Replace the computer cove...

Page 6: ...the NI 1429 and NI 1430 to acquire images from a specific camera Without this camera file the driver does not have the information necessary to configure the NI 1429 and NI 1430 to recognize the image...

Page 7: ...xternal Connections Number of external trigger I O lines 1 Note External TTL lines on the 15 pin D SUB connector become unavailable when the trigger source is set as an optically isolated input or an...

Page 8: ...1 UL 61010 1 CSA 61010 1 Note For UL and other safety considerations refer to the product label or visit ni com certification search by model number or product line and click the appropriate link in t...

Page 9: ...ction for electronic compatibility EMC and product safety You can obtain the DoC for your product by visiting ni com certification If your product supports calibration you can obtain the calibration c...

Page 10: ...l Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the ap...