© National Instruments
|
1-33
STATUS
Yellow
1 flash every
few seconds
Software error—The chassis is unconfigured. Use
MAX to configure the chassis. Refer to the
Measurement & Automation Explorer Help
for
information about configuring the chassis.
2 flashes
every few
seconds
Software error—The chassis has detected an error in
its software. This usually occurs when an attempt to
upgrade the software is interrupted. Reinstall
software on the chassis. Refer to the
Measurement
& Automation Explorer Help
for information about
installing software on the chassis.
3 flashes
every few
seconds
Software error—The chassis is in safe mode
because the SAFE MODE DIP switch is in the ON
position or there is no software installed on the
chassis. Refer to the
information about the SAFE MODE DIP switch.
4 flashes
every few
seconds
Software error—The software has crashed twice
without rebooting or cycling power between
crashes. This usually occurs when the chassis runs
out of memory. Review your RT VI and check the
memory usage. Modify the VI as necessary to solve
the memory usage issue.
Continuously
flashing
Software error—The chassis has detected an
unrecoverable error. Contact National Instruments.
Continuously
flashing or
solid
Software error—The device may be configured for
DHCP but unable to get an IP address because of a
problem with the DHCP server. Check the network
connection and try again. If the problem persists,
contact National Instruments.
Table 1-8.
LED Indications (Continued)
LED
LED
Color
LED State
Indication