NI 9246/9247 Calibration Procedure

5.

Connect the NI cDAQ-9178 chassis to your host computer.

6.

Connect the power source to the NI cDAQ-9178 chassis.

7.

Launch Measurement & Automation Explorer (MAX).

8.

Right-click the device name and select 

Self-Test

 to ensure that the module is working 

properly.

Verification

The following performance verification procedures describe the sequence of operation and test 
points required to verify the NI 9246/9247 and assumes that adequate traceable uncertainties are 
available for the calibration references.

Accuracy Verification

Complete the following procedure to determine the As-Found status of the NI 9246/9247.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9246/9247 to the calibrator as shown in Figure 1.

Figure 1.  

Accuracy Verification Connections

3.

Set the calibrator current to a Test Point Value indicated in Table 4.

4.

Set the calibrator to Operate mode (OPR).

5.

Wait at least 45 seconds for the signal to settle.

6.

Acquire samples from each channel.
a.

Create and configure an AI current task on the NI 9246/9247 according to Table 2.

Table 2.  

NI 9246/9247Configuration for Accuracy Verification 

Min

Max

Scaled Units

Physical Channels

NI 9246

-31.25

31.25

Current

AI0:2

NI 9247

-147

147

Current

AI0:2

NI 9246/9247

C

a

li

b

r

a

tor

HI

LO

AI1

+

AI0

+

AI2

+

Summary of Contents for NI 9246

Page 1: ...tup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Worldwide Support and Services 7 Software Calibrating the NI 9247 requires the install...

Page 2: ...I 9246 specifications and calibration interval NI 9247 Getting Started Guide NI 9247 specific information NI 9247 Datasheet NI 9247 specifications and calibration interval NI DAQmx Readme Operating sy...

Page 3: ...rature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9246 9247 measurement circuitry is at a stable ope...

Page 4: ...curacy Verification Complete the following procedure to determine the As Found status of the NI 9246 9247 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9246 9247 to the calibrator as show...

Page 5: ...only These values only apply under the specific calibration conditions described in this document and are not to be construed as general operational specification of the NI 9246 9247 For operation sp...

Page 6: ...ded calibration currents e Set the calibrator s output frequency to 55 Hz f Set the calibrator to Operate mode OPR g Wait at least 45 seconds for the signal to settle h Call and configure the NI 9246...

Page 7: ...Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Worldwide Support and Services The NI website is your complete...

Reviews: