NI 9246/9247 Calibration Procedure

|

© National Instruments

|

5

b.

Configure the AI current task according to Table 3.

c.

Start the task.

d.

For the maximum Test Point calculate and record the AC magnitude for each channel 
on the NI 9246/9247. NI recommends using the Extract Single Tone Information VI 
to calculate the magnitude. For the minimum test point average the value and record 
the result.

Note

The output magnitude of the Extract Single Tone Information is in Apk. To 

convert to RMS, use the following equation:

e.

Stop and clear the task.

7.

Compare the average for each channel to the limits in Table 4.

Note

The test limits listed in Table 4 should be used for calibration verification 

only. These values only apply under the specific calibration conditions described in 
this document, and are not to be construed as general operational specification of the 
NI 9246/9247. For operation specification, refer to the most recent 

NI 9246 

Datasheet

 and 

NI 9247 Datasheet

 at ni.com/manual. 

8.

Set the calibrator to Standby mode (STBY).

9.

Repeat steps through 7 for each test point listed in Table 4.

10. Disconnect the calibrator from the device.

Table 3.  

NI 9246/9247 Timing Configuration

Sample Mode

Samples Per Channel

Sample Rate (kS/s)

Finite Samples

50,000

50

Table 4.  

NI 9246/9247 Verification Test Limits for Accuracy

Test Point

1-Year Limits

Location

Value 

Frequency 

(Hz)

Lower Limit 

Upper Limit

NI 9246

Max

2 Arms

55

1.9913 Arms

2.0087 Arms

Min

0.00 Adc

0.00

-0.023 Adc

0.023 Adc

NI 9247

Max

2 Arms

55

1.9925 Arms

2.0075 Arms

Min

0.00 Adc

0.00

-0.133 Adc

0.133 Adc

Arms

Apk

(

)

2

(

)

÷

=

Summary of Contents for NI 9246

Page 1: ...tup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Worldwide Support and Services 7 Software Calibrating the NI 9247 requires the install...

Page 2: ...I 9246 specifications and calibration interval NI 9247 Getting Started Guide NI 9247 specific information NI 9247 Datasheet NI 9247 specifications and calibration interval NI DAQmx Readme Operating sy...

Page 3: ...rature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9246 9247 measurement circuitry is at a stable ope...

Page 4: ...curacy Verification Complete the following procedure to determine the As Found status of the NI 9246 9247 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9246 9247 to the calibrator as show...

Page 5: ...only These values only apply under the specific calibration conditions described in this document and are not to be construed as general operational specification of the NI 9246 9247 For operation sp...

Page 6: ...ded calibration currents e Set the calibrator s output frequency to 55 Hz f Set the calibrator to Operate mode OPR g Wait at least 45 seconds for the signal to settle h Call and configure the NI 9246...

Page 7: ...Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Worldwide Support and Services The NI website is your complete...

Reviews: