NI 9246/9247 Calibration Procedure

Documentation

Consult the following documents for information about the NI 9246/9247, NI-DAQmx, and 
your application software. All documents are available on 

ni.com

 and help files install with the 

software.

NI cDAQ-9174/9178 USB Chassis Quick Start

NI-DAQmx installation and hardware setup.

NI 9246 Getting Started Guide 

NI 9246 specific information.

NI 9246 Datasheet

NI 9246 specifications and calibration interval.

NI 9247 Getting Started Guide 

NI 9247 specific information.

NI 9247 Datasheet

NI 9247 specifications and calibration interval.

NI DAQmx Readme

Operating system and application software support.

LabVIEW Help

LabVIEW programming concepts and reference information about NI-DAQmx 
VIs and functions.

NI-DAQmx C Reference Help

Reference information for NI-DAQmx C functions and NI-DAQmx C properties.

NI-DAQmx .NET Help Support for Visual Studio

Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET 
properties, key concepts, and a C enum to .NET enum mapping table.

Summary of Contents for NI 9246

Page 1: ...tup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Worldwide Support and Services 7 Software Calibrating the NI 9247 requires the install...

Page 2: ...I 9246 specifications and calibration interval NI 9247 Getting Started Guide NI 9247 specific information NI 9247 Datasheet NI 9247 specifications and calibration interval NI DAQmx Readme Operating sy...

Page 3: ...rature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9246 9247 measurement circuitry is at a stable ope...

Page 4: ...curacy Verification Complete the following procedure to determine the As Found status of the NI 9246 9247 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9246 9247 to the calibrator as show...

Page 5: ...only These values only apply under the specific calibration conditions described in this document and are not to be construed as general operational specification of the NI 9246 9247 For operation sp...

Page 6: ...ded calibration currents e Set the calibrator s output frequency to 55 Hz f Set the calibrator to Operate mode OPR g Wait at least 45 seconds for the signal to settle h Call and configure the NI 9246...

Page 7: ...Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Worldwide Support and Services The NI website is your complete...