NI 9246/9247 Calibration Procedure

|

© National Instruments

|

3

Test Equipment

Table lists the equipment recommended for the performance verification and adjustment 
procedures. If the recommended equipment is not available, select a substitute using the 
requirements listed in Table 1.

Test Conditions

The following setup and environmental conditions are required to ensure the NI 9246/9247

 

meets published specifications.

Keep connections to the device as short as possible. Long cables and wires act as antennas, 
picking up extra noise that can affect measurements.

Verify that all connections to the device are secure.

Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to 
eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ± 5 °C. The device temperature will be greater 
than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the NI 9246/9247 measurement 
circuitry is at a stable operating temperature.

Initial Setup

Complete the following steps to set up the NI 9246/9247.
1.

Install NI-DAQmx.

2.

Make sure the NI cDAQ-9178 power source is not connected to the chassis.

3.

Connect the NI cDAQ-9178 chassis to the system safety ground.
a.

Attach a ring lug to a 14 AWG (1.6 mm) wire.

b.

Connect the ring lug to the ground terminal on the side of the NI cDAQ-9178 chassis 
using the ground screw.

c.

Attach the other end of the wire to the system safety ground.

4.

Install the module in slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of the 
NI cDAQ-9178 chassis empty.

Table 1.  

Recommended Equipment

Equipment

Recommended 

Model

Parameter 

Measured

Requirements

Source

Fluke 5522A

Accuracy, 

Adjustment

Use a high-precision current 
source with an accuracy 

650 ppm

cDAQ Chassis

NI cDAQ-9178

All

Summary of Contents for NI 9246

Page 1: ...tup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Worldwide Support and Services 7 Software Calibrating the NI 9247 requires the install...

Page 2: ...I 9246 specifications and calibration interval NI 9247 Getting Started Guide NI 9247 specific information NI 9247 Datasheet NI 9247 specifications and calibration interval NI DAQmx Readme Operating sy...

Page 3: ...rature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9246 9247 measurement circuitry is at a stable ope...

Page 4: ...curacy Verification Complete the following procedure to determine the As Found status of the NI 9246 9247 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9246 9247 to the calibrator as show...

Page 5: ...only These values only apply under the specific calibration conditions described in this document and are not to be construed as general operational specification of the NI 9246 9247 For operation sp...

Page 6: ...ded calibration currents e Set the calibrator s output frequency to 55 Hz f Set the calibrator to Operate mode OPR g Wait at least 45 seconds for the signal to settle h Call and configure the NI 9246...

Page 7: ...Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Worldwide Support and Services The NI website is your complete...

Reviews: