NI 9246/9247 Calibration Procedure

Adjustments

The following performance adjustment procedure describes the sequence of operation required 
to adjust the NI 9246/9247.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9246/9247.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9246/9247 to the calibrator as shown in Figure 1.

3.

Adjust the NI 9246/9247.
a.

Initialize a calibration session on the NI 9246/9247. The default password is 

NI

.

b.

Input the external temperature in degrees Celsius.

c.

Call the NI 9246/9247 DAQmx Get C Series Calibration Adjustment Points function 
to obtain an array of recommended calibration current for the NI 9246/9247.

d.

Set the calibrator to a reference value determined by the array of recommended 
calibration currents.

e.

Set the calibrator’s output frequency to 55 Hz.

f.

Set the calibrator to Operate mode (OPR).

g.

Wait at least 45 seconds for the signal to settle.

h.

Call and configure the NI 9246/9247 adjustment function according to Table 5.

Note

To adjust all channels in parallel, use ‘ai0:2’ as the channel name.

i.

Set the calibrator to Standby mode (STBY).

j.

Close and commit the calibration session.

4.

Disconnect the NI 9246/9247 from the calibrator.

EEPROM Update

When an adjustment procedure is completed, the NI 9246/9247

 

internal calibration memory 

(EEPROM) is immediately updated.

If you do not want to perform an adjustment, you can update the calibration date and onboard 
calibration temperature without making any adjustments by initializing an external calibration

setting the C Series calibration temperature, and closing the external calibration.

Table 5.  

Adjustment Configuration

Physical Channel

Reference Value

cDAQMod8/ai

x

The reference value from the array of adjustment points

Summary of Contents for NI 9246

Page 1: ...tup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Worldwide Support and Services 7 Software Calibrating the NI 9247 requires the install...

Page 2: ...I 9246 specifications and calibration interval NI 9247 Getting Started Guide NI 9247 specific information NI 9247 Datasheet NI 9247 specifications and calibration interval NI DAQmx Readme Operating sy...

Page 3: ...rature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9246 9247 measurement circuitry is at a stable ope...

Page 4: ...curacy Verification Complete the following procedure to determine the As Found status of the NI 9246 9247 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9246 9247 to the calibrator as show...

Page 5: ...only These values only apply under the specific calibration conditions described in this document and are not to be construed as general operational specification of the NI 9246 9247 For operation sp...

Page 6: ...ded calibration currents e Set the calibrator s output frequency to 55 Hz f Set the calibrator to Operate mode OPR g Wait at least 45 seconds for the signal to settle h Call and configure the NI 9246...

Page 7: ...Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Worldwide Support and Services The NI website is your complete...