Chapter 1
Getting Started
X Series User Manual
1-4
ni.com
Ferrite Installation
(NI USB-636
x
Mass Termination Devices)
To ens
u
re the specified EMC
performance for radiated RF emissions of the NI USB-636
x
Mass
Termination device, install the incl
u
ded snap-on ferrite bead onto the
power cable, as shown in Fig
u
re 1-2.
Ens
u
re that the ferrite bead is as close to the end of the power cable as
practical. Install the snap-on ferrite bead by opening the ho
u
sing and
looping the power cable once thro
u
gh the center of the ferrite. Close the
ferrite bead
u
ntil the locking tabs engage sec
u
rely.
Yo
u
can order additional EMI s
u
ppression ferrites, 10.2 mm length
(part n
u
mber 781233-02) from NI.
Figure 1-2.
Installing a Ferrite on an NI USB-636
x
Mass Termination Device
1
Power Cable
2
Ferrite
3
NI USB-636
x
Mass Termination Device
x1
3
2
1