10.1 Post-Processing
⇒
7.5.1 POST-PROCESSING on page 47.
This menu contains settings for post-processing the measurement data of broadband HF
probes. These setting options are thus limited to the operating modes of
FIELD
STRENGTH
,
SPECTRUM
and
TIMER LOGGING
. Two types of post-processing are
available:
• The application of a correction factor
• The conversion to display in % of the standard for measured field strengths at a
reference frequency
For evaluating probes, only the correction factor can be applied.
10.1.1 Applying a correction factor
10.1.2 Setting the reference frequency
⇒
7.5.1 POST-PROCESSING on page 47:
Apply Correction Factor CF
When calibrating the HF probes, the measuring error at the defined calibration frequen-
cies is recorded and stored in the probe. If the correction is applied (
Apply Correction
Factor CF
=
On
), then the measuring error is taken into account in the measurement
result and thus a higher measurement accuracy is achieved. However, the correction can
only be applied meaningfully when measuring at a single prevailing frequency, which
must also be known. If this frequency does not exactly correspond to a calibration fre-
quency, it can still be entered as a reference frequency (
Reference Frequency
) and the
corresponding correction factor is then determined by interpolation from the adjacent cor-
rection factors.
⇒
7.5.1 POST-PROCESSING on page 47:
Reference Frequency
Reference Frequency
provides a selection list of probe-specific calibration frequencies.
In addition, the input of any frequency is possible with a resolution of up to 1 kHz.
The last set value in MHz is always displayed, even if the frequency no longer matches
the probe after a probe change. Only when the calibration factor or the display in % of the
standard are activated is the frequency list updated by reading from the probe data and
set to the lowest value.
Summary of Contents for FieldMan
Page 1: ...FieldMan Measurement Device for Electromagnetic Fields User manual...
Page 6: ...6 FieldMan Narda...
Page 10: ...1 About this product 10 FieldMan Narda...
Page 14: ...2 General safety instructions 14 FieldMan Narda...
Page 30: ...5 Controls and display elements 30 FieldMan Narda 5 1 Controls 2 3 4 5 6 1...
Page 90: ...8 Operating modes 90 FieldMan Narda...
Page 108: ...10 Cross mode functions 108 FieldMan Narda...
Page 112: ...11 PC software and updates 112 FieldMan Narda...
Page 120: ...13 Conformity 120 FieldMan Narda...
Page 125: ......