Bed Watcher User’s Guide 

 

 

30.06.2016 

 

Reset  Device  -  forces  the  device  to  be  rebooted.  This  button  does  not  help  when  the  device  is  in 
UltraScope mode.  
 
Set Bootloader - forces the device to stop normal operation and enter in bootloader mode for updating the 
firmware. Please read section 7 of this User's Guide for details about updating the firmware. 
 
Check FW Update - forces the device to stop normal operation and poll specified IP address for firmware 
updates. The utility will offer the user to perform the update when a new firmware file is found. 
 
Get Range - reads maximum device operating range in cm. The device is not capable to detect an object 
beyond this range. 
 
Set Range - programs the device maximum operating range in cm. The device is not capable to detect an 
object beyond this range. The value is taken from the edit box below the Set Range button. 
 
Operating  Range  Criteria  -  selects  the  criterion  which  defines  operating  range.  Sensitivity  defines  the 
range  for  breathing  detection.  Visibility  defines  the  range  for  motion  detection.  Sensitivity  range  is 
always shorter than visibility range.  
 
Get DSP Params - reads the device settings that affect on Alarm, Presence and Motion Flags appearance.  
 
Set  DSP  Params  -  programs  the  device  settings  that  affect  on  Alarm,  Presence  and  Motion  Flags 
appearance. 
 
The following DSP Params can be read or programmed:    
 

 

Sampling Rate, Hz - read-only value - specifies  

 

Noise Threshold, units – specifies sensitivity of the device. The smaller value of the threshold – 
the higher sensitivity has  the device in  terms  of  live object  detection. On the other hand, setting 
the  threshold  too  low  will  make  the  device  over  sensitive  and  the  device  will  be  detecting  live 
objects even in empty premises. Default value is 100 000 energy units. 

 

Motion  Threshold,  units  –  specifies  device  reaction  on  strong  movements.  The  smaller  value  of 
the threshold – the higher probability that the device sets Motion Flag when the object is detected. 
Default value is 10 000 000 000 energy units. 

 

Alarm On Delay, s. – determines degree of confidence for the alarm event. The bigger “Alarm On 
Delay”  is  –  the  higher  confidence  of  the  alarm  event.  On  the  other  hand,  setting  “Alarm  On 
Delay” too big will delay reporting moment about empty premises from the real event further in 
time. Default value is 10 seconds. 

 

Alarm  Off  Delay,  s.  –  reduces  probability  of  false  live  object  detection.  The  bigger  “Alarm  Off 
Delay”  is  –  the  lower  probability  of  the  false  live  object  detection.  On  the  other  hand,  setting 
“Alarm Off Delay” too big will delay reporting moment about live object detection from the real 
event further in time. Default value is 10 seconds. 

 
Get MAC address - displays MAC address of the device. 
 
Get Country Code - displays country code for Wi-Fi operation. Applicable only to the devices with Wi-Fi 
option enabled. 
 
Get  Network  Mode  -  displays  selected  network  mode,  either  AP  for  Wi-Fi  access  point  or  STA  for 
network station (wired or wireless) or TERMINAL for NO/NC relay contacts. 
 

Summary of Contents for Bed Watcher

Page 1: ... implant into the body or other applications intended to support or sustain life or for any other application in which the failure of the NanoPulse product could create a situation where personal injury or death may occur Should Buyer purchase or use NanoPulse products for any such unintended or unauthorized application Buyer shall indemnify and hold NanoPulse Inc and its officers employees subsid...

Page 2: ...atures and Functions 3 3 Installation 4 4 Interfaces and Connectors 4 5 Operation 5 5 1 Power Up 5 5 2 Turning the Device On and Off 5 5 3 Status LEDs 6 5 4 Reading Device Status 6 6 Configuration Options 6 7 Firmware Update 10 8 Rules of Storage and Operation 10 9 Contact Information 11 ...

Page 3: ... on the model either at license free 5 8 GHz ISM or at license free 6 0 8 5 GHz UWB frequency band The device is faced on Figure 1 Figure 1 Bed Watcher 2 Features and Functions Operational range 0 3 4 0 m 1 0 13 1 ft Central frequency 5 800 MHz or 7 200 MHz Frequency range 150 MHz 5 725 5 875 MHz or 500 MHz 7 000 7 500 MHz Antenna pattern 60 degrees Emission power 0 1 mW Power according to IEEE 80...

Page 4: ...rs The Bed Watcher has limited number of electrical connectors which makes it simple placing the device in the network All electrical connectors are located on the top and bottom sides of the device Thus USB 2 0 type B socket is located on the top side of the device as shown on Figure 3 USB Firmware Update Port NANOPULSE INC 7 0 7 5 z UWB SENSOR GH Model 02 00 NP BW S N 00027 160101 BW Figure 3 To...

Page 5: ...D DC TxRx D DC Pin 8 DC TxRx D DC TxRx D DC Tx data line for transmitting data from the device to the network Rx data line for receiving data by the device from the network TxRx bidirectional data line DC positive power line DC common ground level line NO1 normally opened contact 1 NO2 normally opened contact 2 5 Operation 5 1 Power Up Before running the device it is necessary to connect it to the...

Page 6: ... 4 Reading Device Status Bed Watcher device status can be read via ETHERNET RJ45 socket located on the bottom side of the device When STA or AP network mode is enabled the device operates as a TCP IP server and responds on requests from client applications according to the data exchange protocol described in a separate document When TERMINAL network mode in enabled the alarm status is signaled by ...

Page 7: ...ust be inspected and probably repaired Correct device operation is not guaranteed at the current load greater than 500 mA RF Voltage mV engineering parameter Shall be ignored by the user MCU Temperature ºC shows the processor temperature MCU temperature shall not exceed 85 ºC If it happens there may be an issue in the power circuit and the device must be inspected and probably repaired Correct dev...

Page 8: ...read only value specifies Noise Threshold units specifies sensitivity of the device The smaller value of the threshold the higher sensitivity has the device in terms of live object detection On the other hand setting the threshold too low will make the device over sensitive and the device will be detecting live objects even in empty premises Default value is 100 000 energy units Motion Threshold u...

Page 9: ...ess mask assigned to the device either by DHCP server or by the device on its own the default one Get Assigned IP Gate displays IP gateway assigned to the device either by DHCP server or by the device on its own the default one Set Country Code programs in the device country code for Wi Fi operation from the dropdown list Applicable only to the devices with Wi Fi option enabled Set Network Mode pr...

Page 10: ...evice manufacturer or re seller Selecting invalid file will make the firmware update process impossible 6 Update button shall be pressed After that the utility will program the content of the selected firmware file into the device In case of no errors a message saying about successful operation will be shown Otherwise an error message will be displayed describing a possible reason of failure 7 USB...

Page 11: ...in the device 9 Contact Information If you have questions about Bed Watcher operation problem reports or suggestions for the device or documentation improvements please contact NanoPulse Inc representatives NanoPulse Inc 1 Krasnoselskiy per 3 pom I of 75 107140 Moscow Russia Tel 7 926 782 5120 http www nanopulse ru E mail info nanopulse ru ...