Chapter 7
Using the Web Interface
Trillium Compact User Guide
16889R5
2014-03-04
39
Click
Apply
to save your settings. This setting is retained when the Trillium Compact is
powered off or rebooted.
7.5 Sensor Response
The
Sensor Response
link displays a page for viewing (or exporting to a text file) the factory
stored nominal frequency response data for this model and version of the Trillium
Compact. Users can also import, view, and export a text file of user-created information,
such as seismometer response data created by calibrating the seismometer. The
maximum file size that can be imported is 2559 bytes.Hardware Information
The
Hardware Information
link displays a page that lists detailed factory information, such
as the unit model, serial number, and subcomponent versions and serial numbers. This
information can be exported to a text file (.txt) for record keeping or for transmitting to
Nanometrics for technical support purposes.
7.6 Firmware
Use the Firmware page to view the current version of firmware, reboot (restart) the unit, and
to upgrade it with a new firmware file provided by Nanometrics. When restarting the unit
from the Firmware page, a message stating “Goodbye” appears. To access the home
page of the Trillium Compact after it restarts, wait 20 s and return to the home page using
the IP address.
Use the following steps to upload new firmware:
1. Access the Trillium Compact bootloader by
a) Rebooting the firmware and waiting 10 s (or powering up the seismometer); and
then
b) Immediately accessing the URL of the bootloader at:
• http://<IP address>/firmware/bootloader, where <IP address> is the IP address
used to access the Trillium Compact home page.
2. Use the bootloader Web page to upload the new firmware file provided by
Nanometrics.
There is a 10 s interval to access the bootloader before it starts the seismometer firmware. If
this window is missed, the process of restarting or power cycling must be repeated.