Chapter 5
Configuring Your System
Trillium Compact User Guide
16889R5
2014-03-04
31
5.2 Using a Trident Digitizer with a Trillium Compact
Refer to the user guide for your Trident digitizer for complete instructions on using it with a
seismometer. Nanometrics cable 17334-nM (where
n
is the length of the cable in metres)
can be used to connect a Trillium Compact to a Trident digitizer. See
4 for descriptions of these cables.
Following are instructions for configuring your Trident and NaqsServer to work with a Trillium
Compact seismometer.
1. Match the settings on the
Configuration
tab of Nanometrics UI to those in the following
table.
2. Ensure the NaqsServer Naqs.stn file contains the following information:
[ Sensor ] // predefined sensor - all fields mandatory
TypeName = TrilliumCompact
// name of this prototype - may be same as model
Model = TrilliumCompact
// sensor model name
SensitivityUnits = M/S // units of ground motion: M, M/S or M/S**2
Sensitivity = 0.3e9
// counts per unit of ground motion
SensitivityFreq = 1.0 // frequency at which sensitivity is correct
CalibrationUnits = VOLTS // calibration input units: VOLTS or AMPS
CalCoilResistance = 133000
// calibration coil resistance in ohms
CalCoilConstant = 100
// calibration units per m/s/s
CalEnable = 2
// digital enable signal for calibration
CalRelay = 0 // analog relay for calibration (0 = use channel number)
MassCenterEnable = -1
// digital enable signal for mass centering
MassCenterDuration = -1
// duration of mass centering signal (optional)
CalSource = Trident // gives the source of the Cal signal
Note that the
CalCoilResistance
and
CalCoilConstant
parameters need to be
adjusted according to the selected model of Trillium Compact. The computation of
the nominal system sensitivity of your model is described in
32. The correct values for your model can be found in
Section 9.1 “Frequency Response for 120 s and 20 s Models”
45. The
CalCoilConstant
in V/(m/s²) is simply the inverse of the calibration input sensitivity in
(m/s²)/V.
This section does not refer to Trident 305 digitizers. A Trident 305 is configured though the
NMXbus-enabled device to which it is attached, such as a Taurus or Cygnus. Refer to the
Trident 305 user guide and the user guide for the NMXbus device for configuration details.
Nanometrics UI Configuration Tab
Value
Notes
Section
Setting
Front End
Input Range
40 Vpp
Sensor Control
High Voltage Level
+5 V
Calibration Mode
Voltage (active high)
Line 1 Level
Low
Low is equivalent to XYZ mode
and High is equivalent to UVW
mode.
Line 2 Level
Low
Set to High for auto-levelling
and mass centring.
Line 3 Level
Not used