where SEV is the single letter (I, W, E, or F) indicating the severity of the message and NNN is the
serial number of the RM-4 that generated the log message.
Log Messages
Information
BufferedConsumer
A BufferedConsumer is running
A BufferedConsumer process is receiving packets from a PacketReader
process. The PacketReader is the process which reads the packets from the
serial port. A PacketReader is assigned to each serial port. The message is a
notification that the BufferedConsumer has been started and packets
received from the serial ports are forwarded to the central acquisition
process via the Ethernet port.
HrdMsgProc
Starting TelemetryPacketReader
[port number]
The telemetryPacketReader at the serial port specified by the port number in
the message has been started.
Port80
You are not allowed to log in.
The user name and the password received by the RM-4 from the RM-4 user
interface did not match any of the user name and password pairs configured
in the RM-4. Reset the logon window, type in the correct user name and
password and try to login again.
RequestSubmitConfig
Updated config from remote request.
The RM-4 configuration was updated with the new configuration received
from the RM-4 user interface.
File successfully uploaded.
The RM-4 successfully received and accepted the file sent by the RM-4 user
interface.
RM4
Nanometrics RM4 version
[version number],
ROM
[ROM ID].
Informational message about the operating software on the RM-4. The
message is generated startup, and indicates the version number of the
software and the ROM ID (A or B) from which the code is loaded. Operators
should always search for this message while performing the daily system
operation procedures since this message indicates that the RM-4 is
restarted.
Using defaults saved in flash memory.
The RM-4 is using the configuration from the flash memory. Note that the
configuration is written to the flash memory when a “Commit Config”
command is issued by the RM-4 user interface.
RM-4 Manual 44
Appendix D - Instrument Log Messages
Summary of Contents for RM-4
Page 1: ...RM 4 Bridge Multiplexer Manual ...
Page 2: ......
Page 3: ...RM 4 Bridge Multiplexer User Manual ...
Page 8: ...This page intentionally left blank ...
Page 10: ...This page intentionally left blank ...
Page 12: ...This page intentionally left blank ...
Page 40: ...This page intentionally left blank ...