EVE™ PLUS, Automated Cell Counter
21
Operation
Cell counting
(Manual focus)
Optimize the image for analysis such that:
Live cells have bright centers and dark edges.
Dead cells have a uniform blue color throughout the cell with no bright centers.
<Incorrect Image 2>
Dead cells have bright centers are counted as the live cells.
<Incorrect Image 1>
Live cells have dark centers and are counted as the dead cells.
<Correct Image>