26
DATA
Edit the
histogram
1. Choose result on the lists, then press HISTOGRAM button to check detail.
2. Determine the value for minimum cell size and maximum cell size using
up and down arrow button.
» Minimum cell size is used to determine the low range of cell size to in-
clude in the measurement. The algorithm first identifies all objects, and
calculates the average size. From the percent of average size setting, the
algorithm calculates the smallest object size to include in the final measure-
ment. Adjusting the number up, increases inclusiveness thereby decreasing
the lower cell size range.
» Maximum cell size is used to determine the high range of cell size to in-
clude in the measurement. The algorithm first identifies all objects, then
calculates the average size. From the percent of average size setting, the
algorithm calculates the largest object size to include in the final measure-
ment.
3. After modifying cell size, click APPLY button to make the changes.
4. Press BACK button to exit the screen.
5. You can check modified cell size graph and result will be apply on data
lists.