16
+24V
0V
The Side of unit
13
10
2
5
0V
R
②Input/Output Signals
●Output Signal
(PIN No. 2-10, 5-13, 6-14)
There are 3 kinds of output signals: Motor Signal Connect Detector , AUTO MODE and Unit
Power Source Monitoring .
These signals are pulsed transistor activation signals.
Voltage and Current Specifications
Applied Voltage (Vmax)≦30VDC
Working Current (Ip) ≦ 100mA (Rotational Pulse 50mA)
Use an external power source for output circuits. It is recommended to use a separate power from
the one used for Input/Output Connector A. Please refer to
for connections.
R
R
0V
14
6
R
(3)Input/Output Signal Connector Specifications
Input/Output Connector A
Plug Part Number :XM2A-2501 OMRON(or other similar high-quality product)
Cover Part Number:XM2S-2511 OMRON(or other similar high-quality product)
Input/Output Connector B
Plug Part Number :XM2A-1501 OMRON(or other similar high-quality product)
Cover Part Number:XM2S-1511 OMRON(or other similar high-quality product)
・The Plug and Cover are not provided with the system. Please purchase the specified plug and cover
from local suppliers.
・Use only shielded cables to minimize RF interference and noise. Connect the shield to the plug cover.
・Different makers use different names for the cover.
!