16
ALLUX
TM
User’s Guide
【Specifications for backup battery pac】
■
Service temperature: 0℃/+32℉ to +40℃/+104℉
■
Resistance to ambient temperature: -20℃/-4℉ to +60℃/+140℉
■
Resistance to ambient humidity: 10 to 90%RH
■
Battery charger: NE-BC01
■
Applicable device: ALLUX
TM
(NE-Z41 or NE-Z41SH)
■
Service life: 1 year
■
When the vibration indicating the battery exhaustion or requiring the safe
posture sounds, the user must stand in a safe posture immediately.
Even if the user is not in safe posture after 1 minute of warning vibration,
ALLUX
TM
will be forcibly locked or free even while the user is walking and
may result in a fall.
NE-SB01
6.2.4 Battery exhaustion (zero battery)
When the internal power supply has exhausted and ALLUX
TM
cannot be used, the warning
about battery exhaustion will be given by vibration. Then, the vibration for request for safe
posture will be given. When the user takes the safe posture, the walking control will be
stopped, and ALLUX
TM
will be locked (yielding resistance with high flexion resistance) or free
based on choice during adjustment. In this case, ALLUX
TM
can be used continuously for one
day by connecting the backup battery pac (NE-SB01). (For the specifications for the backup
battery pac, see 6.2.3)
ALLUX
TM
will be recovered from this state by charging.
WARNING
States of ALLUX
TM
6
6
States of ALLUX
TM
Summary of Contents for ALLUX NE-Z41