21
ALLUX
TM
User’s Guide
The internal power supply of ALLUX
TM
and backup battery pac shall be charged with
the special battery charger (NE-BC01). When using ALLUX
TM
, it is recommended to
charge it every day as a rule.
■
Remove the prosthesis before charging.
■
If the vibration does not sound when the battery charger is disconnected,
discontinue use, and contact the prosthesis manufacturer. The vibration may
be defective. Cautions and warnings cannot be given. Flexion/extension
resistance may suddenly change, leading to a fall.
■
Do not use for any purpose other than charging ALLUX
TM
.
■
Do not connect any adapter other than the supplied AC adapter to the
battery charger.
■
Do not use the AC adapter for any other electronic devices.
■
Do not charge outdoors.
■
Do not drop from a height. Do not give a strong shock.
■
Do not leave or store at a temperature of -10℃/+14℉ or less or +60℃
/+140℉ or more.
■
Charge indoors in a temperature range from 0℃/+32℉ to +40℃/+104℉.
■
Do not pull the connector cable with a force of 30 N (3 kgf) or more.
■
Do not charge from a car (cigar plug cord).
■
After charging, do not use without the Charging port cap (NE-CC01).
■
Do not put any conductive body, such as metallic parts, to the connector
terminal of ALLUX
TM
.
Failing to follow the above could cause the battery charger to malfunction or
result in the risk of fire.
■
Do not disassemble or modify.
Could result in a malfunction.
■
Do not pour any liquid, such as water.
■
Do not touch with wet hands.
Could lead to a battery short and the risk of electric shock.
7.1.1 Specifications for battery charger and AC adapter
■
Service temperature: 0℃/+32℉ to +40℃/+104℉
■
Resistance to ambient temperature: -10℃/+14℉ to +60℃/+140℉
■
Resistance to ambient humidity: 10 to 90%RH (non condensing)
■
Input voltage: 12 V
■
Input current: Max. 1 A
■
AC adapter: NE-AD01 (Input voltage: 100 V to 240 V)
7.1 Charging ALLUX
TM
7
Charging Procedures
WARNING
Charging Procedures
7
Summary of Contents for ALLUX NE-Z41
Page 43: ...MEMO ...