18
ALLUX
TM
User’s Guide
If a malfunction occurs in a sensor in ALLUX
TM
and it cannot serve as a knee joint, the vibration
for warning about malfunction will sound, and ALLUX
TM
will be locked (yielding resistance with
high flexion resistance). No vibration may sound, or ALLUX
TM
may not be locked depending on
the malfunctioning part.
■
If ALLUX
TM
is defective, the user cannot walk normally.
■
If the vibration for warning about malfunction sounds,
immediately discontinue the use, and contact your prosthetist.
Normal walking will not be supported, possibly resulting in a fall.
【When the temperature is increasing】
4
sec *
2
times with
30
sec interval
* Stop walking with the yielding function or take a rest and stay still until ALLUX
TM
cools
down.
【When the temperature has abnormally increased】
4
sec *
4
times+
10
sec
●
Take the safe posture.
●
ALLUX
TM
will be locked in the flexed state (at a high yielding resistance).
【When ALLUX
TM
cools down】
0.5
*
3
times ×
15
times
:Request for safe posture
⇒ALLUX
TM
will be restored to the normal state even if the user is not in the safe posture
after the completion of vibration for 1 minute.
●
The vibration for confirmation of battery capacity will sound.
0.25
sec×
8
times
:3hours or more
0.5
sec×
8
times
:Less than 3hours
* When ALLUX
TM
is released from the high-temperature state, it has not cooled down
completely. Therefore, if the user continues to walk with the yielding function in the
same manner, ALLUX
TM
will be immediately locked by the high temperature. The user is
recommended to take a rest for 10 to 20 minutes until ALLUX
TM
sufficiently cools down.
WARNING
0.25
sec *
50
times+
5
sec
●
ALLUX
TM
will be locked (at a high yielding resistance).
6.3 In case of malfunction
States of ALLUX
TM
6
6
States of ALLUX
TM