23
PROBLEM
CORRECTIVE ACTION
REFERENCE
Excessive drift of
the rear panel
output voltage.
1.
Check the probe support fixture to
ensure the probe, and the fixture are
not loose.
2.
Check the shield on the sensor cable
and the connectors in the sensor
cable signal path for 100% shield
integrity and connector cleanliness.
1.
PROBE AMPLIFIER CALIBRATION
CHECK for gain adjustment.
2.
PROBE MOUNTING manual section.