7

The  Accumeasure

TM

  System  measures  the  electrical  impedance  of  the  capacitance  between  a  sensing

electrode in the probe and a ground-referenced target.  The magnitude of the impedance is proportional to
the reciprocal of the capacitance value as defined by the equation below:

C

Z

c

1

where:

is proportional to the frequency at which the capacitance measurement is performed.

Substituting the equation for capacitance into the impedance equation shows that the impedance is directly
proportional to the gap value D, as shown in the following equation:

A

D

Z

C



The Accumeasure

TM

 probe amplifier produces a DC voltage that is linearly proportional to the average value

of the probe gap impedance, and an AC voltage variation from the DC voltage that is directly proportional to
the amplitude of the target vibration.  The amplifier electronic circuitry eliminates the effects of both the probe
cable  capacitance  and  the  stray  capacitance  at  the  edge  of  the  probe  sensing  area  that  could  cause  non-
linearity of the gap and vibration measurements.

The  standard  probe  cable  length  is  2.5m  (8.2ft)  and  standard  full  scale  gap  probes  from  25um  to  5mm
(0.001 to 0.200 inches) are available.  Custom cable lengths from 300mm to 15m (1 to 50ft) and custom
probes  from  12.5um  to  12.7mm  (0.0005  to  0.5  inches) can  be  used  with  the  Accumeasure

TM

  probe

amplifier.  Contact the factory for details on custom cables and/or probe lengths.

Summary of Contents for AccuMeasure 9000

Page 1: ...013 AccuMeasure 9000 Capacitance Sensor Amplifier Users Manual MTI Instruments Incorporated 325 Washington Avenue Extension Albany New York 12205 USA Phone 518 218 2550 or 800 342 2203 USA Only Fax 51...

Page 2: ...This page was intentionally left blank...

Page 3: ...i Revision History Rev 4 0 Release...

Page 4: ...vided and upon the expiration of the warranty period of all such a liability shall terminate The foregoing shall constitute the exclusive remedy of the purchaser and the exclusive liability of the Sel...

Page 5: ...al is written in an official language of the Member State Contact MTI Instruments for the availability of translated manuals WARNING ELECTRICAL SHOCK ELECTROCUTION HAZARD This equipment is connected t...

Page 6: ...RN SHIPMENT PROCEDURE 5 2 OPERATING PRINCIPLES 6 3 INSTALLATION INSTRUCTIONS 8 3 1 POWER REQUIREMENTS 8 3 2 CONNECTIONS 9 3 3 CONTROL SWITCHES 13 3 4 DISPLAYS 13 3 5 PROBE MOUNTING 14 3 6 PROBE CABLES...

Page 7: ...parasitic capacitance and ensure the highest linearity Front panel gain and offset controls permit minor adjustments of the probe calibration characteristics An internal 5 pole Butterworth low pass f...

Page 8: ...to 40 C 40 to 100 F Storage Temperature 15 to 65 C 5 to 150 F Dimensions LxWxH 241x165x56mm 9 5x6 5x2 2 Mass 2 3kg 5lbs Probe Voltage 8 5Vrms Max Bargraph Proportional to gap 10 of range per bar 1 Al...

Page 9: ...Cryogenic and Vacuum compatible probes are available Contact the factory Accuracy 0 05 of range when calibrated to a known standard Probe Cable Interchangeability Accurate within 0 5 of range without...

Page 10: ...r optional digital panel meter s should be illuminated Without a probe cable s attached to the probe connector on the front panel all of the bargraph display segments should be fully illuminated Instr...

Page 11: ...al shipping carton if available and forward prepaid to MTI Instruments Inc Supervisor of Manufacturing 325 Washington Avenue Extension Albany NY 12205 5505 Mark RMA and the RMA number if a number is i...

Page 12: ...on the principle of parallel plate capacitor measurement The electrical capacitance formed between an Accumeasure TM probe and a target surface varies as a function of the distance gap between these...

Page 13: ...C voltage that is linearly proportional to the average value of the probe gap impedance and an AC voltage variation from the DC voltage that is directly proportional to the amplitude of the target vib...

Page 14: ...1 Remove the power cord from the rear panel power receptacle 2 Pry up one end of the power receptacle cover by inserting the blade of a small screwdriver under the tab next to the CORCOM lettering 3...

Page 15: ...ding post labeled GROUND on the rear panel see Figure 7 Accumeasure 9000 1 Rear Panel or The displacement voltage output signal s is available from the BNC connector s labeled 1 2 and SUM on the rear...

Page 16: ...fset Switch Indicator Offset Adjustment Probe Input Offset Switch Indicator Offset Adjustment Digital Display Power Indicator Gain Adjustment Figure 3 Accumeasure 9000 1 Front Panel Digital Display Fi...

Page 17: ...2 Gain Adjustment Channel 1 Bargraph Display Channel 1 Offset Switch Indicator Channel 2 Offset Switch Indicator Probe Input Channel 1 Probe Input Channel 2 Channel 2 Offset Adjustment Channel 2 Digi...

Page 18: ...ignal Output 20 W 50 60 Hz 1 100 240 VAC T 0 4A 250V Signal Output Signal Output Ground Connection Power Cable Connection Fuse On Off Switch Signal Output 2 1 20 W T 0 4A 250V 100 240 VAC 50 60 Hz SUM...

Page 19: ...see 4 1 Probe Amplifier Calibration Check The external offset potentiometer is dis engaged when the switch is set to its out position The control is engaged for offset adjustment when the switch set...

Page 20: ...e probe face parallel to the target surface Probe tilt causes measurement errors so a tilt of less than 5 of range is recommended That would mean that for a 500um 20 mil range sensor the edge of the s...

Page 21: ...manufactured feed through assemblies 3 7 Target Grounding The target being measured by the probes connected to the Accumeasure 9000 must be grounded directly to the board power supply grounds for the...

Page 22: ...nge Gain is adjusted at 100 of range Use the following procedure 1 Set the voltmeter to monitor the output of the amplifier 2 With the unit turned off adjust the probe so it is lightly touching the ta...

Page 23: ...n air gap equal to the maximum probe range value for the probe provided with the system The full range value is a function of the probe used and the zero voltage gap value For example a 20 mil range p...

Page 24: ...ush that can be achieved and this is dependant on several factors including probe type and cable length The maximum push recommended is 40X When employing pushed amplifier and probe combinations the o...

Page 25: ...0 volts 8 Connect the DVM or oscilloscope to the Channel 2 output and repeat steps 6 and 7 for Channel 2 output 9 Connect the DVM or oscilloscope to the sum signal output on the rear panel It should...

Page 26: ...00 it should be returned to the factory for servicing Before following the troubleshooting procedure 1 Set up a probe and grounded target so that probe to target gap is within the probe full scale ran...

Page 27: ...ield to center short circuit and probe for interelectrode short circuits 2 If jumper changes on the probe amplifier board have been performed since the unit has been received from MTI contact MTI to r...

Page 28: ...ighted 1 Check the probe cable connection to the front panel ensure that the cable is securely attached to the receptacle 2 Check for an open circuit in the center wire of the probe cable 3 Check for...

Page 29: ...probe support fixture to ensure the probe and the fixture are not loose 2 Check the shield on the sensor cable and the connectors in the sensor cable signal path for 100 shield integrity and connector...

Page 30: ...FROM MTI INSTRUMENTS 1510A Battery Powered Signal Simulator Generator Proforma 300 Semiconductor Wafer Qualification and General Thickness measurement Capacitance MicroTrak Laser Triangulation Displa...

Page 31: ...This page was intentionally left blank...