Command #177: Write Upper Trim Point
Writes the ULTIMA X Upper Trim or Span point value. The ULTIMA Gas Monitor
uses the Upper trim point value to perform Span calibration. When a Span calibra-
tion is performed, the device automatically sets the highest reading obtained to this
Span value. The adjustment range of the Upper Trim Point is from one display unit
to the full scale limit.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0-3
Float
Upper Trim Point [Span] Value
Byte
Format
Description
0-3
Float
Upper Trim Point [Span] Value
Code
Class
Description
0
Success
No Command-specific errors
1 - 2
Undefined
3
Error
Parameter too large
4
Error
Parameter too small
5
Error
Too few data bytes
6
Undefined
7
Error
In write protect mode
8 - 15
Undefined
16
Error
Access restricted
17 - 31
Undefined
32
Error
Busy
33 - 127
Undefined