Appendix: HART Specific Information
MSA
ULTIMA X ® Series
GB
105
Command #129: Read Sensor Gas Type
Reads the Gas Type of the sensor currently connected to the ULTIMA Gas Monitor.
Request Data Bytes
None.
Response Data Bytes
Command #130: Read Device Real Time Clock
Reads the Real Time clock hours and minutes from the ULTIMA X Gas Monitor.
Request Data Bytes
None.
Response Data Bytes
Command #131: Read Alarm Setpoints
Reads the ULTIMA X Alarm Setpoint values.
Request Data Bytes
None.
Response Data Bytes
188
Write Relay Normal State
Byte
Format
Description
0-3
ASCII
Sensor gas type description
[
Table 13.8]
Byte
Format
Description
0
Unsigned
RTC Hours
1
Unsigned
RTC Minutes
Byte
Format
Description
0-3
Float
Alarm 1 Setpoint Value
4-7
Float
Alarm 2 Setpoint Value
8-11
Float
Alarm 3 Setpoint Value
Command #
Description