Ultima MOS-5 Detector
52
8.9.17 CH2 Total Receive Errors (70H)
A read indicates the total Modbus communication hardware receive errors that occurred in the
slave device. The maximum count is 65535 and then the counter will rollover to zero and begin
counting again. The total errors are an accumulation of the individual communication errors
listed below.
8.9.18 CH2 Bus Activity Rate % (71H)
A read indicates the bus activity rate in percent of this slave’s addressed node versus other
addressed nodes. Range of this value is in hex (0-64), which translates to decimal (0-100%).
8.9.19 CH2 Function Code Errors (72H)
A read indicates the number of function code errors that occurred in the slave device. The
maximum count is 65535 and then the counter will rollover to zero and begin counting again.
8.9.20 CH2 Starting Address Errors (73H)
The counter is incremented for each address that does not equal the device address. A read
indicates the number of starting address errors that occurred in the slave device. The maximum
count is 65535 and then the counter will rollover to zero and begin counting again.
8.9.21 CH2 Number of Register Errors (74H)
A read indicates the number of register errors that occurred in the slave device. The maximum
count is 65535 and then the counter will rollover to zero and begin counting again.
8.9.22 CH2 RXD CRC Errors Hi (75H)
A read indicates the number of RXD CRC errors that occurred in the slave device. The
maximum count is 65535 and then the counter will rollover to zero and begin counting again.
8.9.23 CH2 RXD CRC Errors Lo (Same as Hi) (76H)
NOTE:
Hi and Lo CRC errors are now reported in the same word. A read from either Hi or Lo
will return the same count.
8.9.24 CH2 Parity Errors (77H)
A read indicates the number of hardware UART parity errors that occurred in the slave device.
The maximum count is 65535 and then the counter will rollover to zero and begin counting
again.
8.9.25 CH2 Overrun Errors (78H)
A read indicates the number of hardware UART overrun errors that occurred in the slave
device. The maximum count is 65535 and then the counter will rollover to zero and begin
counting again.