![MSA Ultima MOS-5 Instruction Manual Download Page 46](http://html.mh-extra.com/html/msa/ultima-mos-5/ultima-mos-5_instruction-manual_1846250046.webp)
38
Ultima MOS-5 Detector
8.7 Ultima MOS-5 Command Register Locations
Parameter
Function
Type
Scale
Access
Register
Address
Master I/O
Address
Analog
0-20 mA Current Output
Value
16-Bit
R
0000
40001
Mode
Indicates and Controls Mode
Bit
R/W
0001
40002
Status/Error Indicates
Errors
Bit
R
0002 40003
Not Used
N/A
0003
40004
Unit Type
Identifies the Ultima MOS-5
Intelligent Sensor in Decimal
Value 16-Bit
R
0004 40005
Software Rev
Indicates the Software Revision
ASCII
2-Char
R
0005
40006
Status Block
Returns Analog, Mode, Status,
Error, and Sensor Life Returns
Address 6,7,8
Multi 6-bytes
R 0006 40007
Analog Analog
2
Value
R
0006
40007
Mode Mode
Bit
R
0006
40007
Error 2
Error 2
Bit
R
0007
40008
Error 1
Error 1
Bit
R
0007
40008
Sensor Life
Sensor Life
Value
R
0008
40009
Display
Display (LED & MSD)
Bit
/ASCII
R
0009
40010
Display
Display (Mid & LSD)
ASCII
R
000A
40011
Serial Number
Upper serial number
Value
R
000B
40012
Serial Number
Lower serial number
32 bit
R
000C
40013
Alarm Settings Read or change settings for the
high alarm
Bit (0-15)
R/W
000D
40014
Warn Settings Read or change settings for the
low alarm
Bit
(0-15)
R/W
000E
40015
Com1 Addr.
Read or change settings for the
Com1 Address
Value 8-Bit R/W 000F 40016
Com1 Baud
Read or change settings for the
Com1 Baud Rate
Bit
(0-7)
R/W
0010
40017
Com1 Data
Format
Read or change settings for the
Com1 Data Format
Bit (0-7)
R/W
0011
40018
Com2 Addr.
Read or change settings for the
Com2 Address
Value
8-Bit
R/W
0012
40019
Com2 Baud
Read or change settings for the
Com2 Baud Rate
Bit (0-7)
R/W
0013
40020
Com2 Data
Format
Read or change settings for the
Com2 Data Format
Bit
(0-7)
R/W
0014
40021
Not Used
N/A
0015
40022
Reset Alarms
Reset any latched alarms
Bit
(0)
W
0016
40023
Sensor Life
Read the Remaining Sensor
Life
Bit (0)
R 0017
40024
Sensor Scale
Change the Scale for the H
2
S
sensors on the Ultima MOS-5
Intelligent Sensor
Value
8-Bit
R/W
0018
40025
MODEL 10K
Controller (Co)
Indicates Calibration Success
Value
8-Bit
R/W
0019
40026
Not Used
001A
40027
Not Used
001B
40028