11
5.2 Cleaning, Drying and Treating of Components
All components that have been in contact with the mouth or that have been
exposed to the exhalation air of the wearer must be cleaned and disinfected.
These are the following::
−
Breathing tube with mouthpiece, inhalation valve and nose clip
−
Connecting piece with check valve
−
Protective cover
−
Canister
−
Breathing bag with exhalation valve
All other components need only be cleaned if necessary. For the wet cleaning it
is recommended to use the Disinfectant AUER 90 according to instructions for
use. Drying of the components can be at room temperature or in an air stream
of max. 60°C.
After drying and prior to re-assembly talcum powder is spread lightly on the
breathing bag (inside), the heat protector, the breathing tube (outside) and the
protective cover.
Attention: Shake off excess talcum powder!
No talcum powder inside the breathing tube!
5.3 Check of Components
Prior to reassembly all components must be checked to assure they are com-
plete and not damaged. In case components are missing or damaged they
must be replaced. Only original spare parts from MSA AUER GmbH according
to the Spare Parts List shall be used..
5.4 Reassembly
Fig. 12 Push canister into protec-
tive cover all the way.